EN
NIBP cuff socket is at the top of the device.
Note
The connection of the NIBP external air pipe is as shown:
①
Cuff extension tube metal nozzle
②
Socket for air pipe
Figure 1.4.1 The Top External airway Figure 1.4.2 Bottom
At the bottom is the socket for USB:
① The Socket for USB, connect the data line to upload data.
1.5 Accessories
1)
Cuff for adult
2)
USB data line
3)
Disc (PC software)
4)
BP extending tube
5)
Pouch
Note
The monitor also can be equipped with pediatric and neonatal cuff, if necessary, please contact our
company or its representatives.
The width of the cuff should be 40% of the limb circumference (50% for the newborn) or 2/3 of the
length of the upper arm. The length of the inflated part of the cuff should be sufficient to surround
50% to 80% of the limb. Unsuitable cuffs can produce erroneous readings. If there is a problem with
the size of the cuff, use a larger cuff to reduce the error.
Reusable Cuff of adult/ pediatric/ neonatal
Patient
type
Limb
circumference
Width of the
cuff
Length of
inflatable tube
Neonatal
10
~
19 cm
8 cm
1.5 m or 3 m
Pediatric
18
~
26 cm
10.6 cm
Adult 1
25
~
35 cm
14 cm
Adult 2
33
~
47 cm
17 cm
Disposable cuff of neonatal
Size
Limb
circumference
Width of the
cuff
Length of
inflatable tube
1
3.1
~
5.7 cm
2.5 cm
1.5 m or 3 m
2
4.3
~
8.0 cm
3.2 cm
3
5.8
~
10.9 cm
4.3 cm
4
7.1
~
13.1 cm
5.1 cm
Warning
Please use the special accessories supplied by the manufacturer or replace the accessories according
to the requirements of the manufacturer in order to avoid harm to patients.
Note
The monitor also can be equipped with pediatric and neonatal cuff, if necessary, please contact our
company or its representatives.
The cuff is a consumable. In order to correctly measure blood pressure, please replace the cuff in
time.