page 58
cos105000-2016-00 en
Option Description
Comment / Display
OP20 RS232 interface protocol: COM 1
Do not connect any devices, accessories or
software, not listed in “accessories / compatible
devices”.
indicates flashing: number of the RS232 interface protocol
and
reports:
OFF =
RS232 not active / no protocol / interface deactivated
1 =
h/p/cosmos coscom v1, v2, v3 with baud rate 9600 bps
(standard setting COM 1 and COM 2)
3 =
printer protocol (serial printer or converter required)
7 =
TM treadmill emulation in km/h
if available use: h/p/cosmos coscom (= 1)
8 =
TM treadmill emulation in miles per hour
if available use: h/p/cosmos coscom (= 1)
10 =
Loop Back Test (special test plug required, available
at h/p/cosmos)
11 =
SunTech Tango blood pressure monitor
(signal tunnel/loop only)
12 =
Remote Control Hardware Terminal MCU 4
(special hardware required)
20 =
h/p/cosmos coscom v3 with baud rate 115200 bps
For advanced h/p/cosmos coscom v3 connections with baud rate
115200, please select OP20=20. Please note that the connected
devices/software must be approved for h/p/cosmos coscom v3 with
baud rate 115200 (for example h/p/cosmos para control
®
4.1).
OP21 RS232 interface protocol: COM 2
See descriptions above
OP23 RS232 interface protocol COM 4
indicates flashing: Number of the RS232 interface protocol
and
reports:
OFF = RS232 not active / no protocol / interface deactivated
18 = Chip-card reader PROXOMED (special hardware required)
20 = h/p/cosmos coscom v3 / baud rate 115200 bps
22 = POLAR W:I:N:D: - System (more adjustments in user option
OP19 und administrator option OP16)
23 = Chip-card reader ProMedPlus (special hardware required)#
OP27 Minimum acceleration and deceleration level
The selected minimum level is valid for all
acceleration and deceleration processes in all
modes and profiles.
flashes, reports the min. acceleration / deceleration for all
modes and profiles (standard: level 1)
Level of settings: 1 ... 5 but not higher than the value in option 28. For
safety reasons, the acceleration / deceleration levels 5, 6 and 7 cannot
be selected. Note: The selected acceleration and deceleration level is
NOT valid for control and operation via RS232-interface. In this case
the acceleration and deceleration level is set in option 29 or in the
corresponding command of the h/p/cosmos coscom protocol.
OP28 Maximum acceleration and deceleration level
The selected maximum level is valid for all
acceleration and deceleration processes in all
modes and profiles
flashes, reports the max. acceleration / deceleration for all
modes and profiles (standard: level 4)
The maximum acceleration and deceleration level is NOT valid
for control and operation via V24/RS232 interface. In this case
the acceleration and deceleration level is set in option 29 or in the
corresponding command of the h/p/cosmos coscom protocol.