page 54
cos105000-2016-00 en
17 Annex I
17.1 Installation
This device must only be unpacked, transported and installed by h/p/cosmos service personnel (see maintenance).
If the packaging has been damaged, please contact
immediately.
It is the customer’s responsibility to ensure the following conditions before the installation:
❚
There must be a separate power circuit for the device (dedicated line).
❚
There must be a separate wall socket for the device (electrically interlocked with circuit breaker 16A, type C).
❚
The wall socket has to be marked with the serial number of the device.
❚
The wall socket has to be accessible at all times.
❚
The intended location must meet the environmental conditions (see “technical data”).
❚
The intended location must be capable of bearing the load of the device (see “technical data”).
❚
The intended location must provide the safety area and free area as stated in “position of patient and operator”.
❚
The intended location must provide a ceiling height, high enough for accessories (fall prevention).
❚
The intended location must provide a stable and levelled base in order to prevent noise or bouncing.
❚
The intended location must maintain the local requirements for electrical installation.
❚
The intended location must provide a suitable protective earth condition (e.g. PE-bolt).
❚
The intended location must provide the requirements for electrical installation acc. to “technical data”.
The manufacturer does not assume liability for any damage, complaints or missing parts that are not reported immediately upon
delivery on the packing list/delivery note.
17.2 Instruction protocol
When installing the device, the h/p/cosmos service personnel instructs the intended operator according to the instructions for use,
following this instruction protocol.
With the name and signature on the instruction protocol, the instructed persons confirm that they fully know how to operate the
device safely. The instructed persons confirm they are able to instruct further operators according to this protocol.
No.
Information
Chapter
Check
1.
These instructions for use are available as print version at
.
-
2.
The instructions for use are to be read in full before starting with the operation.
-
3.
The safety information is explained in detail.
Safety
4.
The safety information must be displayed within sight of the device.
Safety
5.
The function of all safety equipment is explained in detail.
Safety
6.
The use of the prescribed fall prevention is explained in detail.
Safety
7.
The position of patient and operator is explained.
Position of P+O
8.
The safety area (2 m behind device) is specifically pointed out especially.
Position of P+O
9.
The function of the UserTerminal is explained in detail.
UserTerminal
10.
The general use is explained (incl. manual, profile, cardio and test mode).
Operation
11.
The competences and intervals for maintenance are explained.
Maintenance
12.
The adjustment of the running belt is explained in detail.
Maintenance
13.
The lubrication of the device is explained in detail.
Maintenance
14.
The accompanying documents are explained and handed over.
-