13.12 Voltage, Current, Performance
Description
Data
Input voltage *
200 V - 240 V ~ (f: 50 – 60 Hz)
Current input (long time) *
6.0 A
Current input (momentary) *
13.5 A
Energy consumption (long time)
≤ 1440 VA
Energy consumption (momentary)
≤ 3240 VA
Energy efficiency
N/A
Device protection switch (circuit breaker)
16 A
Drive motor capacity
2200 W
Elevation motor capacity
470 W
Earth leakage current
≤ 0.2 mA
Isolation transformer
1840 VA
Power supply cord
detachable, 3 m
13.13 All-pole disconnection
There are the following options for all-pole disconnection:
❚
Unplug device from power socket.
❚
Unplug cable from device.
❚
Switch off device protection switch
Regard free area in order to maintain access to cables and circuit breaker (see “position of patient and operator”).
* Overload or weak power supply may lead to reduced speed accuracy or tripping the fuse.