User Manual
sens07vest
Page 12
May 2020
2.4
Replacing actuator module & CO2 cylinder
The actuator module needs to be replaced after each inflation.
Disassembly of the sens07 inflator
1
Remove CO
2
cylinder by firmly rotating cylinder counter clock-
wise.
2
Remove sens07 head by rotating rippled screw nut counter
clockwise.
3
Remove actuator module from the sens07 head and discard.
Rearming of the sens07 inflator
1
Check the date (month & year) of the sens07 head. The date
should not be over today’s date.
2
Insert a new and unused actuator module into the sens07
head.
3
Install the sens07 head by screwing rippled screw nut clock-
wise until it meets the housing shoulder.
4
Install an unused CO
2
cylinder by rotating clockwise into base
unit until CO
2
cylinder is secured firmly.
Perform the self-test as described after replacing the actuator
module.
User Manual
sens07vest
May 2020
Page 13
3
LED signals
The inflator head has 2 front LED for important information and 3
back LED for system information.
3.1
Main LEDs on the front side
The green and red front LED signal the following information:
3.1.1
Water zone indicator
A short blink every second indicates water detection. The number
of blinks corresponds to the dive zone.
The red LED is used instead of the green LED in case of a negative
self-test. However, the sens07 inflator will still try to operate as
expected, but there is a high chance of failure.
The water zone indicator can be disabled for defense applications.
3.1.2
Suspend Mode
An alternating green & red LED blink means that the button was
pressed under water suspending inflation for a certain time.
sens07 inflator
is functional
sens07 inflator
has a fault
Safe zone
• • • •
• • • •
Diving zone
•• •• •• ••
•• •• •• ••
Warning zone
••• ••• ••• •••
••• ••• ••• •••
sens07 inflator
is functional
Suspend mode
• • • • •
• • • •