3
CP - OPTI
Technical Specifications
X-ray film processorss (MDD class I) are employed in “medical” applications (med-
ical products directive) and “non-medical” applications (electrical appliances and
EMC directives).
Intended use includes observing the operating instructions, the installation instruc-
tions and adherence to the safety notes.
Any application differing from intended use voids the guarantee by PROTEC
®
.
The owner of the machine will be liable for damages resulting from unintended use
or faulty application.
Intended use includes adherence to all statutory regulations concerning occupa-
tional safety and radiation protection applicable at the operating site.
Technical Specifications
Film transport:
Continuous roller transport system
Film formats:
In general: Sheet and roll films up to 35.8 cm (14.1 ’’)
width;
Roll films with leader from 70 mm (2.8 ’’) width;
Smallest film format 10x10 cm (4x4 ’’).
Mammography type 1171: For processing mammog-
raphy films.
-
-
Processing capactiy:
1
11
films 24x30 cm (10x12 “) per hours (standard
model, film fed in crosswise)
Process time:
Standard
105
s
Mammography 135 s
(
*Option
90 s /
167 s
)
Linear speed:
Standard
48
cm/min. (
19
in/min)
Mammography 37 cm/min (14.5 in/min)
*Option
56 cm/min (22 in/min) /
30 cm/min (11.8 in/min)
Developer time:
Standard
29
s Mammography 37 s *Special
25 s /
46 s
Tank capacities:
Developer, fixer
and
washing 5 litres each (1.3 gal)
Circulation system:
Developer and fixer are continiuously circulated by a
circulation pump
Replenishment:
Automatic replenishment by film detection, in relation
to film length
Developer temperature:
Adjustable 28 - 37 °C (82.4 - 98.6 °F)
Fixer temperature:
Adjusted to developer temperature by heat
exchanger.
Water connection:
Permissible water pressure 2 - 10 bar (29 - 145 psi),
permissible water temperature 5 - 30 °C (41 - 86 °F).
Water consumption:
1.9 litres per minute (0.5 gal/min) when processing.
Drain capacity:
7 litres per minute (1.85 gal/min).
Noise level:
Less than 58 dB(A).
Heat emission:
Stand-by: 0.1 kJ/s Processing: 1.4 kJ/s
Environmental condi-
tions:
1) Temperature 18 - 40 °C (51.6 - 104 °F), venti-
lated room, room temperature should be lower
than set bath temperature.
2) Relative humidity lower than 80% up to 31 °C
(80 °F), linear decreasing to 50% at 40°C (104°F)
3) Height above sea level less than 2000 m (6666 ft.)