3
Now by pressing one of the buttons 4 or 5 change the display: “L.0” if you want to unlock the
input or “L.1” if you want to lock the input.
4
To finalize the changes press button 6 “Exit”.
NOTE
Even when the input lock is activated it is possible to recall (previously stored)
processing parameters of the memory-function.
CAUTION!
Please note, if input is locked via lock function, then processing parameters
cannot be stored - when trying to do this, an error message “LO” will be
displayed.
5.2.16
Stop Film Transport
In a case of a film-jam inside the machine, the film transportation can be manually interrupted. To stop
the film transport press both arrow-buttons (2 and 3) simultaneously.
Related topics:
See “Manual starting and stopping the film transport:” 5.2.13.3
See “Film is caught up in the racks”: see 5.4.2.10
5.2.17
Infrared Sensitive Materials
Infra-red sensitive film material can be processed without danger of exposure by the light-barrier.
5.2.18
Automatic Cooling Function
If the developer temperature exceeds the set value, the automatic cooling function will be activated.
The fan runs for a fast cooling of the bath. The same happens when the temperature is decreased by