COMPACT 2 NDT
User Manual
5190-0-0002
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
26 von 65
5.2.10
Dimmer Function of Display
With the dimmer function the intensity of the control panel display can be changed. The intensity can
be reduced to avoid possible exposure of high-sensitive films. Normal-sensitive films are not affected
by the display.
5.2.10.1
Setting the intensity of the display
1
Switch processor on
2
Press arrow „down „button (3) for about 5 seconds. The intensity changes. Two intensity-levels
are available, bright and dark. During film-throughput this functions is not available
5.2.11
Machine cover Safety Disconnection
The machine cover may only be removed for service and maintenance purposes. The processor cannot
be started without the cover. In the event that the machine cover is removed during film-throughput,
the film transport will be stopped. On the display the error message
“E1”
will be displayed. The error will
be reset when the machine cover has been replaced.
5.2.12
Auto fill Function
In case new chemicals have to be filled into the processor (after installation, tank cleaning), the tanks
can be filled automatically by means of the autofill function. In the process, the tank is filled for a fixed
period of 17 minutes, that is, chemicals are pumped from the replenishment containers to the tanks.
Also the water bath will be filled (8 min. period). The display will show two symbolized tanks. When the
autofill function has been completed, the machine enters the stand-by mode. In the event of the
respective baths being full before the time is up, the autofill function can be manually stopped.
5.2.12.1
Starting up the auto fill function
1
The chemicals containers need to be filled manually with at least 1.0 litres of fresh chemicals, to
ensure the pumps will not run dry!
2
Switch processor off
3
Filling developer
and
fixer tank: Keep both replenishment buttons for developer (7) and fixer (8)
pressed down and switch the machine on.
or filling
developer
or
fixer tank: Keep respective
replenishment button (7 or 8) pressed down and switch the machine on.
5.2.12.2
Manual cancellation of the auto fill function
1
Stop developer tank filling: Keep developer replenishment button (7) pressed down and press
“arrow down” button (3).
2
Stop fixer tank filling: Keep fixer replenishment button (8) pressed down and press “arrow
down” button (3).
The filling of the water bath will be stopped once the automatic filling of both chemicals baths has
been stopped.
5.2.13
Manual Mode
In the manual mode, the processor works without the light barrier. The film transport has to be started
and stopped manually. All the set values in the manual mode are also valid in the automatic mode.
Please note that the Infeeding-film-display (“– –”) is deactivated. In the Manual Mode the replenishment
function does not use film surface measurement, instead it uses a standard value. In the manual mode,
a replenishment cycle can also be activated manually.
5.2.13.1
Switching to manual mode:
Switch the machine on. During stand-by, press the arrow-buttons “up” (2) and “down” (3)
simultaneously. When in manual mode the display is flashing.
5.2.13.2
Switching back to automatic mode:
In manual mode with film transport off press both arrow-buttons “up” (2) and “down” (3)
simultaneously.