COMPACT 2 NDT
User Manual
5190-0-0002
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
27 von 65
5.2.13.3
Manual starting and stopping the film transport
1
Switch to manual mode
2
Press the processing time button (4) – the button illuminates
3
Start the film transport by pressing the arrow-button "up" (2) or stop the film transport by
pressing the arrow-button „down“ (3)
5.2.13.4
Manual replenishment
1
Switch to manual mode
2
Press required replenishment button for developer (7) or fixer (8) – respective button
illuminates
3
Start the replenishment cycle by pressing the arrow-button „up“ (2) or cancel the
replenishment cycle by pressing the arrow-button „down“ (3)
5.2.14
Use of Memory Function
In the memory five sets of parameters can be stored and be recalled into working memory.
5.2.14.1
Store processing parameters
1
Set machine to preferable parameters e.g. bath temperature, processing time etc. (see manual).
2
Press buttons 4 and 5 “Enter Memory” simultaneously to enter the memory mode.
3
Use buttons 2 and 3 “Select” to select the parameter memory (P1 to P5). Inside this the
parameters will be stored, old values will be overwritten.
4
Press button 5 “Store” to store parameters and to leave the memory mode.
5.2.14.2
Recall processing parameters
1
Press buttons 4 and 5 “Enter Memory” simultaneously to enter memory mode.
2
Use buttons 2 and 3 “Select” to select the parameter memory (P1 to P5) from which the
parameters shall be recalled.
3
Press button 4 “Recall” to recall parameters (copy into working memory) and to leave the
memory mode.
5.2.14.3
To leave memory mode without change
Press button 6 “Exit”.
CAUTION!
Please note, if input is locked (via lock function), then processing parameters
cannot be stored - when trying to do this, an error message “LO” will be
displayed. Despite locked input, the processing parameters can be recalled.
CAUTION!
Choose the same bath temperature for all stored programmes. Of course
different bath temperatures can be stored too, but when changing the
programme you always have to wait until the changed bath temperature is
reached.
5.2.15
Lock Function
To prevent unintentional change of operation parameters the input can be locked. If the input is locked
any attempt to change the parameters will display an error “LO” (locked).
5.2.15.1
Activating and de-activating of the input lock
1
Press buttons 4 and 5 “Enter Memory” simultaneously to enter the memory mode (see diagram
see 5.2.15).
2
Press several times the button 3 “Select” until the display will show “L.0”(Lock off input
unlocked) or “L.1” (Lock on: input locked).