Polydoros RFX
Instructions for use
5021-0-1002
19 von 64
WARNING
Accessible voltages.
Contact with accessible voltages may cause serious physical injury or death.
To avoid the risk of electric shock, this equipment must be disconnected from supply before
opening.
WARNING
Accessible voltages.
Contact with accessible voltages may cause serious physical injury or death.
Before any work within the opened generator or at the XTA, the generator shall be
disconnected from mains by opening the internal mains circuit breaker.
Remark: the switch off function of the generator does not disconnect the mains voltage from
generator line input. After switch off all components of the generator and the XTA are still
powered. After disconnect from mains, behind fuses F1-F3 all components of the generator
and the XTA are deenergized.
CAUTION
Usage of wrong fuses.
Risk of fire.
Replace fuses only with same type and rating.
CAUTION
Being exposed to X-ray radiation without knowing it.
Risk of moderate injury caused by X-ray radiation.
Only enter the examination room, if the X-ray indicator is off.
CAUTION
Specified load limits of generator are exceeded.
Plastic smolders and produces toxic gases.
Limiting values for usage of the generator must not be exceeded.
WARNING
DC-link is powered.
Contact with line voltage may cause serious physical injury or death.
Remove the MRC cover only, when all LEDs on the MRC-board(s) are off.