5
Error messages
The software CONAXX 2 will display error messages if the generator has an abnormal status. This
chapter contains a table for those messages and suggests actions to be taken.
Error ID
Error message
Action
0x1BE
(446)
The submodule RCE has exceeded
temperature of 70°C
Switch off generator and cool down
0x202
(514)
The operator has opened the door
during exposure
-
0x203
(515)
The XTA has reached maximum
temperature / pressure
Cool down XTA
0x204
(516)
The XTA temperature has reached 70°C
No action necessary
0x210
(528)
Internal SW error: HW selftest wrong
step.
Reset error and restart generator
0x211
(529)
Internal SW error: HW selftest not
allowed.
Reset error and restart generator
0x222
(546)
The door contact is open during selftest
-
0x225
(549)
The requested tube current is too high.
Check exposure settings
0x226
(550)
The requested tube current is too low.
Check exposure settings
0x227
(551)
The requested tube voltage is too high.
Check exposure settings
0x228
(550)
The requested tube voltage is too low.
Check exposure settings
0x246
(582)
The exposure shall not be interrupted by
the user by releasing the exposure
button.
Do not release button during exposure
0xD00
(3328)
Adapter detected implausible exposure
parameters
Check exposure settings
0xD01
(3329)
Internal SW error: Communication error
between Adapter and generator
software
Restart generator
0xD02
(3330)
An error occurred in calculation of dose
compensation
Check exposure settings
NOTE
When other errors appear, please contact a PROTEC authorized service technician.