Polydoros RFX
Instructions for use
5021-0-1002
61 von 64
3.2
Exposure Release
The exposure is released either with the exposure button integrated in the Mini Console or with the
hand switch (optional).
NOTE
Before releasing X-ray, check the selected acquisition data at the workstation. For
X-ray release, the workstation must be in ready state.
Preparations (pre contact):
Exposure button:
Press the exposure button down to the first pressure point and hold it approx. 1 sec
Hand switch:
Press the exposure button down to the first pressure point and hold it approx. 1 sec
•
The rotating anode is accelerated to nominal speed
•
The required filament current is set
NOTE
After warming up the filament and accelerating the rotating anode, it is possible to
release X-ray.
Radiation release (main contact):
Exposure button:
Press the exposure button down fully and keep it pressed until the exposure is ended
Hand switch:
Press the exposure button down fully and keep it pressed until the exposure is ended
NOTE
Releasing the exposure button immediately ends the X-ray.
X-ray indication:
The radiation LED on the control console lights up and a buzzer signal sounds during
exposure release.