Polydoros RFX
Instructions for use
5021-0-1002
24 von 64
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±0.5 kV
±1.0 kV
±2.0 kV
±0.5 kV
±1.0 kV
±2.0 kV
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % U
T
0.5 cycle
at 0°, 45°, 90°,
135°, 180°, 225°,
270°, and 315°
Not applicable for
products with 3-
phase AC mains
and input current
rated with more
than 16 A per
phase.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the product
requires continued operation during
power mains interruptions, it is
recommended that the product be
powered from an uninterruptible
power supply or a battery.
0 % U
T
1 cycle
70 % U
T
25/30
cycles
Single phase at
0°
Voltage
interruptions
IEC 61000-4-11
0 % U
T
:
250/300 cycles
0 % U
T
:
250/300 cycles
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
50/60 HZ
30 A/m
50/60 HZ
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.