background image

Notes:

  

1. These results were obtained with SimCube 

1.5 software. We are constantly improving the 

product so future software versions may give 

slightly different results.

2.  These  results  represent  data  taken  from 

specific  models  of  these  manufacturer's 

products.  We  normally  expect  a  given 

manufacturer's results to be pretty consistent 

across their product line, but different models 

and even different software revisions may give 

different  results.    For  example  an  HP 

OmniCare  with  software  E.0.1  reads  adult 

mode  diastolic  about  10mmHg  lower  than  a 

HP Merlin with software 17.52.

3.  The  readings  shown  are  actual  random 

samples,  but  they  do  not  necessarily  reflect 

worst case.

4. MDE = Escort II SW 4.2.0, HP = Merlin SW 

17.52,  Critikon  =  Vitalnet  2200,  Datascope  = 

Accutorr 3, CIS = 506DXNT

Optimizing Your Simulations

 The following techniques will help you obtain 

the best possible simulation results.

1.  Cuff  Size  Selection.

  Many  vendors  are 

immune to variations in cuff size, but some are 

not. For example we find that some Datascope 

Accutorr  monitors  produce  systolic  values 

about  2mmHg  higher  and  diastolic  values 

about 6mmHg higher with a 400ml cuff volume 

rather than a 200ml cuff volume, while some 

Critikon monitors show about a 6mmHg drop 

in systolic values when changing to the 400ml 

cuff volume.  We recommend using something 

close  to  a  400ml  cuff  volume  for  adult  mode 

simulation.  This can be achieved by wrapping 

an  adult  cuff  on  itself  to  form  a  cylinder 

approximately  3  inches  in  diameter.      If  you 

want to quantify the volume of a rolled cuff you 

can do this by injecting a measured amount of 

air into an unrolled, empty cuff using a syringe 

and  a  check  valve.    Once  you  have  injected 

the desired amount of air, roll the cuff tightly.

SimCube adult mode simulation will normally 

work  with  a  broad  range  of  test  volumes,  but 

with  smaller  test  volumes  you  may  see  more 

vendor to vendor reading differences.

2. Be prepared to throw out the first reading.

We find that some vendors sometimes produce 

a first reading that is substantially different than 

later readings.

3. Take an average of several readings.

 We 

suggest using a minimum of 3 to 5 readings.

4.  Make  sure  there  are  no  leaks.

  You  can 

usually see leaks just by watching the SimCube 

manometer during the reading.  On step down 

systems the cuff pressure may normally drop a 

bit  during  the  first  few  steps,  but  after  a  few 

steps it should be flat or even rise slightly during 

the step.  This is because the air in the cuff is 

cooled by the rapid step deflation and warms up 

again  during  the  step  holding  period  causing 

the pressure to rise slightly.  A leak on a step 

down system will appear as drop in pressure on 

all steps.   Bleed down systems should normally 

bleed at about 2 to 6mmHg per second.  A leak 

on a bleed down system will appear as a faster 

than normal bleed rate.  Since the SimCube's 

NIBP  simulation  creates  pulses  which  are 

several mmHg in size it is easiest to check for 

leaks  with  the  SimCube  in  manometer  mode.  

Be  sure  to  watch  for  leaks  on  the  SimCube's 

manometer  rather  than  the  monitor's  cuff 

pressure display as the SimCube has a much 

faster sample rate and display update rate than 

most NIBP monitors.

5.  Make  sure  the  monitor  is  in  calibration.

On monitors with quickly updating cuff pressure 

displays you can do a rough cal check by simply 

placing  the  SimCube  on  top  of  the  monitor, 

placing the SimCube in manometer mode, and 

starting  a  reading.   The  two  pressure  gauges 

should  track  tightly,  although  the  monitor's 

display may be noticeably delayed.  On some 

monitors, however, the displayed pressure may 

represent  the  intended  pressure  for  the 

pressure step rather than the actual pressure. 

Revision 

O

 0

7/18/

2017 

 www.pronktech.com  

800-541-9802

Page 19

PRONK TECHNOLOGIES

Содержание SimCube

Страница 1: ...SimCube NIBP Simulator Operator s Manual www pronktech com...

Страница 2: ...orming to the relevant standards and directives required for CE have a CE Mark label as shown below affixed to the unit CE Marked products also comply with the WEEE directive for category 9 Monitoring...

Страница 3: ...18 2017 www pronktech com 800 541 9802 Page 3 PRONK TECHNOLOGIES Getting Started 4 NIBP Detailed Information 6 ECG RESP Detailed Information 8 IBP Detailed Information SC 5 Only 9 Battery Operation 1...

Страница 4: ...e are printed on the front panel next to the LED horizontally For more detailed information on each mode refer to each parameter section in this manual 5 For dynamic NIBP testing connect the SimCube s...

Страница 5: ...supplied with the SimCube system and the correct IBP adapter for the monitor under test to the SimCube simulator A variety of IBP adapters are available from Pronk Technologies Inc Select the 0 mmHg...

Страница 6: ...scribed below Refer to the front panel of your SimCube simulator to determine which modes apply to your model NIBP Adult 120 80 Simulates a patient with blood pressure of 120 80mmHg heart rate of 70 b...

Страница 7: ...ses the monitor must be placed in a special test mode so that it will not release the cuff pressure NIBP Leak Test SC 5 only This mode is used for measuring leak rates in NIBP monitors their cuffs and...

Страница 8: ...tion mode SC 5 only The Arrhythmia mode is designed to allow you to quickly check that all the life critical ECG alarms on a patient monitor are properly configured The mode consists of a single cardi...

Страница 9: ...he NIBP values associated with the same mode and are indicated by the arrow on the front panel in the IBP column pointing towards the NIBP column IBP 0 Use this mode to simulate atmospheric pressure a...

Страница 10: ...g features in your new SC 5 as compared to any previous versions you may have The display will be slightly dimmer The SimCube will go into sleep mode when the NIBP functions are not in use the display...

Страница 11: ...that the switch is in the Batt OFF position when you are done the numeric display on the SimCube simulator is blanked under certain conditions to conserve power so you cannot rely on that to indicate...

Страница 12: ...SimCube System Troubleshooting Tips Revision O 07 18 2017 www pronktech com 800 541 9802 Page 12 PRONK TECHNOLOGIES...

Страница 13: ...noise by running the SimCube on battery power to see if the noise still exists If it is present check to see that the monitor under test has its A C filter s enabled Please note that monitors may hav...

Страница 14: ...tors This cable has a six pin mini DIN to GE style connector It is six inches long and works with the IBP Extension cable IBP Interface Cable for Datex Ohmeda Monitors This cable has a six pin mini DI...

Страница 15: ...s an annual calibration interval Extended Warranty up to 5 years total may be purchased only at time of original sale or with purchase of Rejuvenation Service Revision O 07 18 2017 www pronktech com 8...

Страница 16: ...se or use Pronk Technologies shall be liable only for repair or replacement of the SimCube NIBP Simulator and optional features Pronk Technologies shall not be liable for any incidental or consequenti...

Страница 17: ...l readings but 5 might be only within 21mmHg and 1 might be as far out as 29mmHg While these specifications may seem very loose they can be quite difficult to meet The errors in the readings come from...

Страница 18: ...at they should be statistically identical but in practice they can still give different results 2 Every manufacturer has adjusted their system based on data which was measured by the nurses they used...

Страница 19: ...ormally work with a broad range of test volumes but with smaller test volumes you may see more vendor to vendor reading differences 2 Be prepared to throw out the first reading We find that some vendo...

Страница 20: ...h the best setup there is still variation from reading to reading Sometimes a reading will be produced which is quite far off target and it is difficult to know if this represents a problem with the m...

Страница 21: ...Revision O 07 18 2017 www pronktech com 800 541 9802 Page 21 PRONK TECHNOLOGIES...

Страница 22: ...Revision O 07 18 2017 www pronktech com 800 541 9802 Page 22 PRONK TECHNOLOGIES...

Страница 23: ...Revision O 007 18 2017 www pronktech com 800 541 9802 Page 23 PRONK TECHNOLOGIES...

Страница 24: ...erent models and sometimes even different software versions can give quite different readings but the following are some average values obtained from a variety of devices Revision O 007 18 2017 www pr...

Отзывы: