User Interface Information
SC-4 SimCubes with software version 3.3 and SC-5 SimCubes with software version 4.3 and
above include the ‘Press and Hold’ feature. This feature allows the user to advance through
various simulations in a given mode. To place the SimCube into the Press and Hold mode, press
and hold the Yellow
Mode
button for two seconds or until the
Mode
LED begins to flash. Each
press of the Yellow
Mode
button will advance to the next waveform. As an example, when in HR
Seq, by placing the SimCube in the Press and Hold mode, the HR Seq will stay at 30BPM until the
Yellow
Mode
button is pressed. Then it will advance to 60BPM. Each subsequent press of the
Yellow
Mode
button will advance to the next waveform in the sequence. To exit the Press and
Hold mode, simply press and hold the Yellow
Mode
button for two seconds, or power cycle the
SimCube.
NIBP Detailed Information
Before you begin, make sure you have your SimCube setup as described in ‘Getting Started.’
Available simulation modes are described below. Refer to the front panel of your SimCube
simulator to determine which modes apply to your model.
•
NIBP Adult (120/80)
Simulates a patient with blood pressure of 120/80mmHg, heart rate of 70 bpm, mean pressure of
97mmHg, and pulse volume of 1 ml.
•
NIBP Neo (70/40)
Simulates a patient with blood pressure of 70/40mmHg, heart rate of 95 bpm, mean pressure of
51mmHg, and pulse volume of 0.25 ml. Requires use of a neo or pediatric cuff and the monitor
under test should be placed in neonatal mode, if available.
• NIBP High (190/120)
Simulates a patient with blood pressure of 190/120mmHg, heart rate of 70 bpm, mean pressure of
142mmHg, and pulse volume of 1 ml.
• NIBP Low (80/40) (SC-5 only)
Simulates a patient with blood pressure of 80/40mmHg, heart rate of 70 bpm, mean pressure of
58mmHg, and pulse volume of 1 ml.
• NIBP Peak Detect / Over Pressure (SC-3, 4 & 5 only)
This mode is used to test the over pressure feature on patient monitors. The SimCube readout
will latch the highest pressure value sensed (in 1mmHg steps). When the input pressure has
returned to 0 mmHg for 15 seconds, the latched value is cleared. Typically the monitor must be
Revision
O
0
7/18/
2017
www.pronktech.com
800-541-9802
Page 6
PRONK TECHNOLOGIES
Содержание SimCube
Страница 1: ...SimCube NIBP Simulator Operator s Manual www pronktech com...
Страница 21: ...Revision O 07 18 2017 www pronktech com 800 541 9802 Page 21 PRONK TECHNOLOGIES...
Страница 22: ...Revision O 07 18 2017 www pronktech com 800 541 9802 Page 22 PRONK TECHNOLOGIES...
Страница 23: ...Revision O 007 18 2017 www pronktech com 800 541 9802 Page 23 PRONK TECHNOLOGIES...