Pronk Technologies SimCube Скачать руководство пользователя страница 10

Battery Operation

SimCube  models  SC-3,  SC-4  and  SC-5  may  be  operated  on  battery  power  using  the  Battery 
Boost  Option.   The  Battery  Boost  allows  the  user  to  switch  between AC  power  and  four  “AA” 
batteries and it easily integrates with the SimCube padded carrying case. The Battery Boost is 
designed to operate with alkaline batteries for simplicity and convenience, however, please note 
that using NiMH batteries will provide the highest battery performance.  Simulating only NIBP, the 
SimCube  simulator  can  perform  100-200  NIBP  simulation  readings  on  a  set  of  fresh  NiMH 
batteries.  While rechargeable batteries may be used, an external charger is required to recharge 
these batteries.

SC-5 models with software version 5.0 or higher include enhancements to help extend battery life 
while using the Battery Boost Option. Please note the following battery saving features in your new 
SC-5 as compared to any previous versions you may have:

• The display will be slightly dimmer.

• The SimCube will go into "sleep" mode when the NIBP functions are not in use, the display will 

go blank and the green Mode LED will flash.

• When the batteries are depleted, the Battery Level indicator on the Battery Boost will illuminate 

a red LED and the SimCube may re-boot to further indicate the need to replace the batteries. 

1. CRITICAL OPERATION NOTES.  PLEASE READ FIRST!

•  You may use NiMH rechargeable batteries, but the Battery Boost does not charge them.

Rechargeable batteries must be charged outside of the SimCube system.

•  When the Battery Level indicator lamp is red your simulator may continue to operate, but 

values should not be trusted. Always confirm the Battery Level LED is green when doing a 

simulation on battery power.

•  When your SimCube simulator is not in use, switch the Battery Boost to the “Batt OFF” 

position to avoid draining your batteries.

Revision 

O

 0

7/18/

2017 

 www.pronktech.com  

800-541-9802

Page 10

PRONK TECHNOLOGIES

Содержание SimCube

Страница 1: ...SimCube NIBP Simulator Operator s Manual www pronktech com...

Страница 2: ...orming to the relevant standards and directives required for CE have a CE Mark label as shown below affixed to the unit CE Marked products also comply with the WEEE directive for category 9 Monitoring...

Страница 3: ...18 2017 www pronktech com 800 541 9802 Page 3 PRONK TECHNOLOGIES Getting Started 4 NIBP Detailed Information 6 ECG RESP Detailed Information 8 IBP Detailed Information SC 5 Only 9 Battery Operation 1...

Страница 4: ...e are printed on the front panel next to the LED horizontally For more detailed information on each mode refer to each parameter section in this manual 5 For dynamic NIBP testing connect the SimCube s...

Страница 5: ...supplied with the SimCube system and the correct IBP adapter for the monitor under test to the SimCube simulator A variety of IBP adapters are available from Pronk Technologies Inc Select the 0 mmHg...

Страница 6: ...scribed below Refer to the front panel of your SimCube simulator to determine which modes apply to your model NIBP Adult 120 80 Simulates a patient with blood pressure of 120 80mmHg heart rate of 70 b...

Страница 7: ...ses the monitor must be placed in a special test mode so that it will not release the cuff pressure NIBP Leak Test SC 5 only This mode is used for measuring leak rates in NIBP monitors their cuffs and...

Страница 8: ...tion mode SC 5 only The Arrhythmia mode is designed to allow you to quickly check that all the life critical ECG alarms on a patient monitor are properly configured The mode consists of a single cardi...

Страница 9: ...he NIBP values associated with the same mode and are indicated by the arrow on the front panel in the IBP column pointing towards the NIBP column IBP 0 Use this mode to simulate atmospheric pressure a...

Страница 10: ...g features in your new SC 5 as compared to any previous versions you may have The display will be slightly dimmer The SimCube will go into sleep mode when the NIBP functions are not in use the display...

Страница 11: ...that the switch is in the Batt OFF position when you are done the numeric display on the SimCube simulator is blanked under certain conditions to conserve power so you cannot rely on that to indicate...

Страница 12: ...SimCube System Troubleshooting Tips Revision O 07 18 2017 www pronktech com 800 541 9802 Page 12 PRONK TECHNOLOGIES...

Страница 13: ...noise by running the SimCube on battery power to see if the noise still exists If it is present check to see that the monitor under test has its A C filter s enabled Please note that monitors may hav...

Страница 14: ...tors This cable has a six pin mini DIN to GE style connector It is six inches long and works with the IBP Extension cable IBP Interface Cable for Datex Ohmeda Monitors This cable has a six pin mini DI...

Страница 15: ...s an annual calibration interval Extended Warranty up to 5 years total may be purchased only at time of original sale or with purchase of Rejuvenation Service Revision O 07 18 2017 www pronktech com 8...

Страница 16: ...se or use Pronk Technologies shall be liable only for repair or replacement of the SimCube NIBP Simulator and optional features Pronk Technologies shall not be liable for any incidental or consequenti...

Страница 17: ...l readings but 5 might be only within 21mmHg and 1 might be as far out as 29mmHg While these specifications may seem very loose they can be quite difficult to meet The errors in the readings come from...

Страница 18: ...at they should be statistically identical but in practice they can still give different results 2 Every manufacturer has adjusted their system based on data which was measured by the nurses they used...

Страница 19: ...ormally work with a broad range of test volumes but with smaller test volumes you may see more vendor to vendor reading differences 2 Be prepared to throw out the first reading We find that some vendo...

Страница 20: ...h the best setup there is still variation from reading to reading Sometimes a reading will be produced which is quite far off target and it is difficult to know if this represents a problem with the m...

Страница 21: ...Revision O 07 18 2017 www pronktech com 800 541 9802 Page 21 PRONK TECHNOLOGIES...

Страница 22: ...Revision O 07 18 2017 www pronktech com 800 541 9802 Page 22 PRONK TECHNOLOGIES...

Страница 23: ...Revision O 007 18 2017 www pronktech com 800 541 9802 Page 23 PRONK TECHNOLOGIES...

Страница 24: ...erent models and sometimes even different software versions can give quite different readings but the following are some average values obtained from a variety of devices Revision O 007 18 2017 www pr...

Отзывы: