background image

Optimizing Simulation Results With Your SimCube Oscillometric 

NIBP Simulator  

by Karl Ruiter, Pronk Technologies Inc.

Introduction

The  SimCube  NIBP  simulator  provides 

accurate  NIBP  simulation  in  an  amazingly 

convenient,  rugged,  and  affordable  package.  

As  you  use  your  simulator  on  various  patient 

monitors it is important to understand what sort 

of results to expect, what they mean, and how 

to optimize them.

NIBP Device Accuracy

The accuracy of NIBP devices is governed by 

an  AAMI  standard  (SP-10).    This  standard 

specifies two different types of accuracy: static 

and dynamic.  

   The static accuracy represents the accuracy 

with which the internal pressure gauge in the 

device  can  measure  cuff  pressure  and  it  is 

required  to  be  within  +/-3mmHg  or  +/-2%, 

whichever  is  greater.    The  static  accuracy  of 

the  device  can  be  easily  checked  by  placing 

the monitor in cal or check mode and verifying 

the  accuracy  against  the  SimCube's 

manometer mode.

    The  dynamic  accuracy  represents  the 

accuracy of the device in taking actual readings 

on  patients,  and  it  is  specified  in  a  relatively 

vague  way.    The AAMI  specification  calls  for 

automated (device) readings to be taken on a 

large group of patients meeting certain criteria 

and  the  results  to  be  compared  with  manual 

readings taken by nurses on the same patients.  

The manual and automatic readings are then 

compared statistically.

      To  be  acceptable  under  the  standard  the 

mean  error  on  systolic  and  diastolic  readings 

must  be  no  more  than  5mmHg  and  the 

standard  deviation  of  the  error  must  be  no 

more than 8mmHg.

   What this means is that, under the specifica-

tion, roughly 68% of all patients must be within 

13mmHg of the manual readings, but 5% might 

be only within 21mmHg and 1% might be as far 

out  as  29mmHg.  While  these  specifications 

may seem very loose they can be quite difficult 

to meet.

The errors in the readings come from several 

basic sources:

1.Patient  to  Patient  Physiological  Variations:

Different  patients  have  different  arterial  pulse 

shapes,  arterial  compliance,  flesh  rigidity  and 

other  factors  which  simply  make  the  BP  cuff 

respond differently. The oscillometric signal is 

complex  and  changes  not  only  in  size  but  in 

shape with cuff pressure and it is simply slightly 

different from patient to patient. Additionally, a

patient's  actual  BP  values  are  often  not 

perfectly  stable,  but  change  over  time,  often 

during a reading or a series of readings.

 2. Extrinsic Noise:  During the testing process 

patients may be moving or talking.  The signal 

to  noise  ratio  on  the  oscillometric  signal  is 

never very good and it does not take very much 

additional noise to affect readings.

3.  Intrinsic  Noise:    Even  on  a  perfectly  still 

patient  with  perfectly  static  blood  pressure 

there  would  still  be  significant  reading  to 

reading variations.  The biggest factor on this is 

sampling error introduced by the cuff pressure 

bleed  rate  or  step  size.  It  is  no  accident  that 

step-down  NIBP  monitors  step  in  8mmHg 

steps and the standard deviation of the error is 

specified  at  8mmHg.  In  addition  to  this  both 

bleeding  and  stepping  introduce  pneumatic 

noise which can affect the oscillometric signal.

NIBP Device Differences

  In manual NIBP readings there is a published 

AHA standard for how the readings should be 

performed. In practice the process is somewhat 

subjective,  but  at  least  there  is  the  intent  of 

uniformity.

  With automated NIBP readings the situation is 

quite different. The first companies in the   

Revision 

O

 0

7/18/

2017 

 www.pronktech.com  

800-541-9802

Page 17

PRONK TECHNOLOGIES

Содержание SimCube

Страница 1: ...SimCube NIBP Simulator Operator s Manual www pronktech com...

Страница 2: ...orming to the relevant standards and directives required for CE have a CE Mark label as shown below affixed to the unit CE Marked products also comply with the WEEE directive for category 9 Monitoring...

Страница 3: ...18 2017 www pronktech com 800 541 9802 Page 3 PRONK TECHNOLOGIES Getting Started 4 NIBP Detailed Information 6 ECG RESP Detailed Information 8 IBP Detailed Information SC 5 Only 9 Battery Operation 1...

Страница 4: ...e are printed on the front panel next to the LED horizontally For more detailed information on each mode refer to each parameter section in this manual 5 For dynamic NIBP testing connect the SimCube s...

Страница 5: ...supplied with the SimCube system and the correct IBP adapter for the monitor under test to the SimCube simulator A variety of IBP adapters are available from Pronk Technologies Inc Select the 0 mmHg...

Страница 6: ...scribed below Refer to the front panel of your SimCube simulator to determine which modes apply to your model NIBP Adult 120 80 Simulates a patient with blood pressure of 120 80mmHg heart rate of 70 b...

Страница 7: ...ses the monitor must be placed in a special test mode so that it will not release the cuff pressure NIBP Leak Test SC 5 only This mode is used for measuring leak rates in NIBP monitors their cuffs and...

Страница 8: ...tion mode SC 5 only The Arrhythmia mode is designed to allow you to quickly check that all the life critical ECG alarms on a patient monitor are properly configured The mode consists of a single cardi...

Страница 9: ...he NIBP values associated with the same mode and are indicated by the arrow on the front panel in the IBP column pointing towards the NIBP column IBP 0 Use this mode to simulate atmospheric pressure a...

Страница 10: ...g features in your new SC 5 as compared to any previous versions you may have The display will be slightly dimmer The SimCube will go into sleep mode when the NIBP functions are not in use the display...

Страница 11: ...that the switch is in the Batt OFF position when you are done the numeric display on the SimCube simulator is blanked under certain conditions to conserve power so you cannot rely on that to indicate...

Страница 12: ...SimCube System Troubleshooting Tips Revision O 07 18 2017 www pronktech com 800 541 9802 Page 12 PRONK TECHNOLOGIES...

Страница 13: ...noise by running the SimCube on battery power to see if the noise still exists If it is present check to see that the monitor under test has its A C filter s enabled Please note that monitors may hav...

Страница 14: ...tors This cable has a six pin mini DIN to GE style connector It is six inches long and works with the IBP Extension cable IBP Interface Cable for Datex Ohmeda Monitors This cable has a six pin mini DI...

Страница 15: ...s an annual calibration interval Extended Warranty up to 5 years total may be purchased only at time of original sale or with purchase of Rejuvenation Service Revision O 07 18 2017 www pronktech com 8...

Страница 16: ...se or use Pronk Technologies shall be liable only for repair or replacement of the SimCube NIBP Simulator and optional features Pronk Technologies shall not be liable for any incidental or consequenti...

Страница 17: ...l readings but 5 might be only within 21mmHg and 1 might be as far out as 29mmHg While these specifications may seem very loose they can be quite difficult to meet The errors in the readings come from...

Страница 18: ...at they should be statistically identical but in practice they can still give different results 2 Every manufacturer has adjusted their system based on data which was measured by the nurses they used...

Страница 19: ...ormally work with a broad range of test volumes but with smaller test volumes you may see more vendor to vendor reading differences 2 Be prepared to throw out the first reading We find that some vendo...

Страница 20: ...h the best setup there is still variation from reading to reading Sometimes a reading will be produced which is quite far off target and it is difficult to know if this represents a problem with the m...

Страница 21: ...Revision O 07 18 2017 www pronktech com 800 541 9802 Page 21 PRONK TECHNOLOGIES...

Страница 22: ...Revision O 07 18 2017 www pronktech com 800 541 9802 Page 22 PRONK TECHNOLOGIES...

Страница 23: ...Revision O 007 18 2017 www pronktech com 800 541 9802 Page 23 PRONK TECHNOLOGIES...

Страница 24: ...erent models and sometimes even different software versions can give quite different readings but the following are some average values obtained from a variety of devices Revision O 007 18 2017 www pr...

Отзывы: