Figure 7. Orientation of Magnetic Rod Assembly.
The orientation of the magnetic rod assembly, with the
screw holes facing the back of the instrument.
Figure 7 shows the correct orientation of the magnetic rod assembly. The screw holes face the back, and the
magnetic rods sit toward the front of the instrument. If the rod assembly needs to be removed for cleaning
(see Section 8) and replaced, ensure that this orientation is retained.
3.
Hardware Mode
The hardware mode chosen for each run depends on the purification protocol being used. Please refer to the
Technical Bulletin or Manual for the various Maxwell
®
Purification Kits for information on the hardware mode
setting required.
The existing hardware mode (SEV or LEV) setting is displayed in the upper right hand corner of the
Home
screen.
The instrument displays the hardware mode entered but does not sense the hardware installed. Always verify that
the hardware installed matches what is displayed on the
Home
screen. Make sure that the hardware and hard-
ware mode are appropriate for the purification kit being used. If the installed hardware is not correct for the kit
being used, remove the installed magnetic rod assembly and replace it with the appropriate magnetic rod assembly
(see Section 7.A).
If the installed hardware does not match the hardware mode displayed, turn the machine o
ff
, and install the
appropriate hardware with the machine turned o
ff
. When the machine is powered up after the installation, the
installed hardware will match the hardware mode displayed on the
Home
screen.
Back
Front