background image

5

 

© 2009 by Proceq SA

3.  Real Tests 

 

3.6 Accept or modify limits  

C omplete G etting S tarted 

7.8 Pre-define R oll IDs  in 

P aperlink and download to 

hammer 

Method 1: P erform an 

impact (R oll ID = C ounter) 

Method 2: Manually 

enter R oll ID 

Method 3: S elect a pre-

defined R oll ID 

 

3.3 Meas ure the R oll Hardnes s  P rofile 

3.4 T erminate the S eries  and review 

3.1 P reparation 
 

 

3.2 Define a R oll ID for the T es t S eries  

Содержание 342 10 000

Страница 1: ...more than 50 years of know how you can measure Made in Switzerland Operating Instructions Paper Roll Hardness Tester...

Страница 2: ...list 9 4 Technical Specifications 9 5 Part Numbers and Accessories 10 5 1 Units 10 5 2 Parts and Accessories 10 6 Maintenance and Support 10 6 1 Self test function 10 6 2 Checking the calibration on t...

Страница 3: ...r hammer as described in chapter 6 Congratulations Your new PaperSchmidt is fully operational and you can now continue with your measurements Performing a reset In case a reset is required this can be...

Страница 4: ...s damage resulting from the effects of foreign bodies accidents vandalism and force majeure All information contained in this documentation is presented in good faith and believed to be correct Proceq...

Страница 5: ...R oll IDs in Paperlink and download to hammer Method 1 Perform an impact R oll ID Counter Method 2 Manually enter R oll ID Method 3 S elect a pre defined R oll ID 3 3 Meas ure the R oll Hardnes s Pro...

Страница 6: ...Roll ID and will appear in the review list once the series has been terminated Roll ID Counter Method 2 Manual entry of the Roll ID Method 3 Select a pre defined Roll ID Note A single pre defined name...

Страница 7: ...ct if it was an obvious outlier Dustbin icon Scroll to the desired icon and press the SELECT button again A confirmation screen prevents terminating a series pre maturely Use cancel X symbol or confir...

Страница 8: ...will shift left 3 6 Accept or modify limits This function is a rapid way to check if the roll hardness profile is within a pre defined tolerance After selecting a pre defined Roll ID or after enterin...

Страница 9: ...mm Dimensions of housing 55 x 55 x 250 mm 340 mm to tip of plunger Dimensions plunger visible part 105 x 15 mm radius of spherical tip 25mm Weight 570 g Memory Data No of impacts in a series Maximum 9...

Страница 10: ...rized service centre 6 2 Checking the calibration on the anvil The PaperSchmidt checks for correct impact energy with each and every impact Nevertheless it is recommended to check the calibration on a...

Страница 11: ...used Connecting to the PaperSchmidt Connect the PaperSchmidt to a USB port then click on the icon to bring up below window Leave the settings as default or if you know the COM port you can en ter it...

Страница 12: ...t the limits directly in this window by clicking on the blue limit values Note The user may also add a comment to the measurement series To do so click on Add Note The user may alter the order in whic...

Страница 13: ...another program such as Excel If you wish to export the individual impact values of the series you have to display them by clicking on the double arrow icon as described above before you Copy as text...

Страница 14: ...Deleting data stored on the PaperSchmidt Select the menu item Device Delete all Data on Device to delete all data stored on the Paper Schmidt A warning will be given to say that the data is about to...

Страница 15: ...utomatically onto the PaperSchmidt Up to 50 series names can be defined in this way Each name can be up to 12 characters long Note Please note that also copy and paste works with data from Excel 7 9 T...

Страница 16: ...W Shanghai 200050 Tel 86 21 63177479 62122680 62122698 Fax 86 21 63175015 info china proceq com Proceq Rus LLC Ul Kompozitorov 10 lit A pom 345 N 194356 St Petersburg Russia Phone 7 812 335 92 34 Fax...