13
Professional ultrasonic nebulizer
How to prepare therapy
3. Attaching the nebulization unit
Remove the Isapak® chamber
❺
from its sterile
packaging and attach it to the chamber holder
❹
.
Warning: The interior of the Isapak® aerosol chamber
is
❺
sterile and shall not come into contact with a
non-sterile object under any circumstances.
4. Filling a inhalation solution
Two filling options can be selected.
1. Filling the solution directly into the Isapak® chamber
❺
Attach Isapak® adapter
❻
to the sterile water bottle
❼.
Place the Isapak® sterile water bottle
❼
into the
Isapak® chamber
❺
from above. After the aerosol
chamber
❺
has filled with sterile water (50 ml),
remove the Isapak® sterile water bottle
❼
.
Carefully fill the medicine into the Isapak® aerosol
chamber
❺
.
Remove the Isapak® sealing stopper
❽
from its
sterile packaging and use it to close the Isapak® aerosol
chamber
❺
without touching the lower part of the
Isapak® sealing stopper
❽
.