19
Professional ultrasonic nebulizer
How to finish therapy
Warnin
g
When assembling the cleaned, disinfected, and dried parts, do not touch the
places where the medication and nebulized medication pass through directly
with your hands to prevent possible infection.
Do not make any contact to the vibrator while the power plug is plugged into
the electric outlet.
• You may suffer electric shock or injure yourself.
PRODUCT CARE & MAINTENANCE
How to take care of the bacterial filter
Replace the bacterial filter
❷
after each 50 hours of operation or more often if it becomes
dirty (due to the high level of dirt in the air).
CLEANING AND DISINFECTION
Warning: Adhere to the following hygiene instructions at all times to prevent any health
hazard through contaminated PROFI SONIC A.
Clean the dust from the main unit
❶
, inhalation hose holder
⓬
and chamber holder
❹
by wiping it with a disposable cloth after being moistened with water or light-detergent
solution.
General advice
• Do not wipe the main unit with volatile chemicals, such as benzene or thinners.
• Do not damage the surface of the vibrator.
• Please observe local disposal guidelines for used accessory parts such as disposable products.
Some parts of the unit may be damaged if you use disinfectants based on:
phenols,
alkyl amine compounds,
halogen-releasing compounds,
strong organic acids,
oxygen-releasing compounds.