Chapter 8
Dual Pulse Combiner
63
<500
µs
>1
µs
Suggested PG-200 Settings
Pulse Width:
50 nsec to 200 µsec
PG-200 Input 1 Delay:
As dictated by experiment timing; immaterial to Combiner.
PG-200 Input 2 Delay:
The interval from the end of the first pulse to the start of the
second can range from 1 µsec to 500 µsec with a 50 nsec gate width. The delay
set at PG-200 Input 2 must always be short enough to allow the second pulse to
be completed within 500 µsec of the end of the first pulse as indicated in Figure
22.
Figure 22.
Combiner
maximum pulse
separation.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com