Chapter 6
Timing Modes
53
Example 7
Laser (or other experimental equipment)
HV Gate Pulser
Controller
Trigger out
Ext sync
Trigger in
Delay generator
Inhibit
SHUTTER MONITOR (ST-130,
ST-133, ST-138, PentaMAX)
This example illustrates the use of a multichannel delay generator to control both
experiment and WinView/WinSpec timing. One channel of the delay generator triggers
the laser. At the appropriate time, the second channel triggers the pulser. After gating,
the pulser can be used to trigger the controller, which is in the External Sync mode, or
the controller can be left in the Freerun mode and SHUTTER MONITOR used to enable
the pulser.
Time
HV gate pulse
0
CCD readout
Acquire
Readout
0
0
0
Delay generator signal 1
Delay generator signal 2
Not scan
(Shutter in ST-138)
Figure 19.
Use of
multichannel
delay generator
to control
experiment and
WinView/
WinSpec
timing.
Figure 20.
Timing for
experiment
depicted in
Figure 19.