11.4
Performing an automatic caustic measurement
This chapter aims to provide some basic information as you get to know the FocusMonitor FM+, discussing
the example of an automatic caustic measurement with the LaserDiagnosticsSoftware LDS.
For a detailed description of the software installation, file management and evaluation of the measured data,
please refer to the separate operating manual LaserDiagnosticsSoftware LDS.
11.4.1
Selecting the measuring mode automatic caustic
1. Connect the device according to
chapter 11.3 on page 28 with the
LaserDiagnosticsSoftware LDS.
The FocusMonitor FM+ is established
as a connected device.
2. Click on the
Scanner
function.
The
Device control
menu opens.
3. Click on the
Automatic caustic
but-
ton or
on the drop-down list measurement
mode and
Automatic caustic
.
Function Scanner
Drop-down list Measuring mode
The corresponding
Device control
opens.
The
Automatic caustic
tool bench
opens with the following tools:
•
Caustic analysis
•
Plane analysis