25
© 2013 PreSens Precision Sensing GmbH
Fibox 4 & Fibox 4 trace
Operation
1.
Sensor Settings
In the
Sensor Settings
you have to choose the
Sensor type
of your new sensor (Fibox 4:
PSt3; Fibox 4 trace: PSt3, PSt6 or PSt9) and type in the
Sensor Constants
.
!
You can find the sensor constants and calibration data on the Final Inspection
Protocol delivered with your oxygen sensor.
Change the values on the display to the values stated on your sensor´s Final Inspection
Protocol, then move to the upper right of the screen to the
Next
button and press .
2.
Calibration Data
Select
Dry
or
Humid
according to the conditions under which the calibration data was
obtained and press . Then change the calibration values to the values stated on the
sensor´s Final Inspection Protocol:
Cal0 = first calibration point at 0 % oxygen
T0 = temperature at which the first calibration point was measured
Cal2nd = second calibration point
T2nd = temperature at which the second calibration point was measured
O2-2nd = oxygen concentration of the second calibration standard
pATM = atmospheric pressure at which the calibration was performed
!
Please note that for the O2-2nd and the pATM values the respective unit can be
changed as well. Make sure that the selected unit matches the one stated on the Final
Inspection Protocol.
When you have adjusted all the calibration values move to the upper right of the screen to the
Next
button and press .
Fig. 33
Calibration Data screen