Rate Meter Programming
Overview
The meter can also be used to display flow rate. In addition to the scaling and
calibration procedures described above, the only setup required for this is setting
the meter to display rate, and programming the low-flow cutoff if required.
Set Display to Rate (
dSPy r
)
The user may select either process/rate or total to be set as the default displayed
reading. When displaying process/rate, the process/rate LED indicator will be
illuminated.
To change the display from reading total
to rate
Press
ENTER
to begin scrolling through the
functions.
When display rate (
dspy r
)
appears, press
ENTER
.
The meter now displays rate and the green
“
R
” LED on the right side is illuminated.