Model PD6262 Analog Dual-Input Rate/Totalizer Instruction Manual
42
MeterView
®
Pro Software
The meter can also be programmed using the PC-based MeterView Pro
software available for free download at www.predig.com.
Data logging for one meter at a time is available with MeterView Pro
software. More advanced data acquisition may be accomplished by
using any Modbus RTU compliant software.
In order to program the meter using a computer, the meter must be
connected using a USB, RS-232, or RS-485 serial adapter, see
ORDERING INFORMATION on page 7 for details.
Input Calibration Method (
InCAL
)
There are two methods of calibrating (or scaling) the display for each
input channel to show the correct engineering units.
Use
the
Scale
menu to enter the scaling without a signal source.
Use
the
Calibrate
menu to apply a signal from a signal source.
Note: The Scale and Calibrate functions are exclusive of each other.
The meter uses the last function programmed. Only one of these
methods can be employed at a time. The Scale and Calibrate functions
can use up to 32 points (default is 2). The number of points should be
set in Scale and Calibrate accordingly under the Number of Points
(
nopts
)
menu selection prior to scaling and calibration of the meter, see
page 78 for details.