METER OPERATION
The meter is capable of accepting two input channels (A and B) of either
current (0-20 mA, 4-20 mA) or voltage signals (0-5 V, 1-5 V, 0-10 V,
10 V) and displaying these signals in engineering units from -99999 to
999999 (
e.g.
a 4-20 mA signal could be displayed as -50.000 to
50.000). A totalizer can be programmed to count the scaled engineering
units, interpreting it as count per second, minute, hour, or day. The
scaled rate and total for each channel can be displayed on the top or
bottom displays.
A math function channel (C) is available to perform operations on
channel A and B rates or totals, with adder and factor constants, and
display the results. Engineering units or tags may be displayed with
these three channels.
The dual-line display can be customized by the user. Typically the upper
display is used to display the math channel C, while the lower display is
used to alternate between displaying input channels A and B rate or
total.
Additionally the meter can be set up to display any input or math
channel on the upper display and a unit or tag on the lower display. The
relays and analog output can be programmed to operate based on any
input or math channel.
Front Panel Buttons Operation
Button
Symbol
Description
Press to enter or exit Programming Mode,
view settings, or exit max/min readings
Press to reset max/min readings or other
parameter/function assigned through the
User
menu
Press to display max/min readings for channel
A or other parameter/function assigned
through the
User
menu
Press to acknowledge relays or other
parameters/function assigned through the
User
menu