P
RO
V
U
™ PD6310-WM NTEP Certified Pulse Input Batch Controller
Instruction Manual
52
Troubleshooting
Due to the many features and functions of the
controller
, it’s possible that the setup of the controller
does not agree with what an operator expects to see.
If the controller is not working as expected, refer to
the
Diagnostics
menu and recommendations below.
Diagnostics Menu (
diag
)
The
Diagnostics
menu is located in the
; to access the
Diagnostics
menu refer
to page 47.
This menu allows the user to test the functionality of
all the controller LEDs, check the controller
’s software
and version information, and erase the MeterView Pro
software installation files from the controller. Press the
Enter
button to view the settings and the Menu button
to exit at any time.
For a description of the diagnostic messages,
see
Advanced Features Menu & Display Messages, page
47.
Testing the Display LEDs
To test all LEDs on the display:
1. Go to the
Diagnostics
menu (
diAG
) and
press Enter button.
2. Press Up arrow button and scroll to
LED Test
menu (
led t
)
3. Press the Enter button to activate the LED
Test. The controller will cycle through all
digits, decimal points, and relay indicators to
enable the operator to check that all LEDs
are functioning properly.
4. Press the Enter button again to access the
Information
menu (
info
) or press the Menu
button to return to Run Mode.
Determining Software Version
To determine the software version of a controller:
1. Go to the
Diagnostics
menu (
diAG
) and
press Enter button.
2. Press Up arrow button and scroll to
Information menu (
Info
).
3. Press Enter to access the software
number (
SFT
) and version (
UER
)
information. Write down the information as
it is displayed. Continue pressing Enter
until all the information is displayed.
4. The controller returns to Run Mode after
displaying all the settings.
Reset Controller to Factory
Defaults
When the parameters have been changed in a way
that is difficult to determine what’s happening, it might
be better to start the setup process from the factory
defaults.
Instructions to load factory defaults:
1. Enter the
Advanced Features
menu. See
2. Press Up arrow to go to
Diagnostics
(
diag
) menu
3. Press and hold Right arrow for three seconds,
press Enter when display flashes
reset
.
Note: If Enter is not pressed within three
seconds, the display returns to the
Diagnostics
menu.
4. The controller goes through an initialization
sequence (similar as on power-up) and loads the
factory default settings.