SERIES-300 S / SCS / SIP
PAG OERLIKON AG
CALIBRATION INSTRUCTIONS
D 04.06
The calibration of the internal reference weight is concluded
The balance is automatically turnd off and on again
The balance is ready for the final control
The calibration of the internal reference weight can also be carried out with the service box:
- Connect the service box to the balance
- Put the rotary switch of the service box at position "0"
- Press key "9" of the service box
- The calibration cycle starts directly at page D 04.05
The calibration operation of the internal reference weight for gauged balances is described on the next page
Содержание 300 SCS series
Страница 1: ...PAG OERLIKON AG The Balance of Quality SERVICE MANUAL SERIES 300 S 300 SCS 300 SIP...
Страница 4: ...PAG OERLIKON AG SERIES 300 S SCS ELEMENTARY DIAGRAM PRINZIPSCHALTBILD ELEMENTARY DIAGRAM A 01 01...
Страница 121: ...PAG OERLIKON AG SERIES 300 S SCS WEIGHING PAN D 150 WAAGSCHALE D 150 B 06 01 Group M1 M2 2 3 4 5 6 7 8 1...
Страница 123: ...PAG OERLIKON AG SERIES 300 S SCS WEIGHING PAN D 170 WAAGSCHALE D 170 B 07 01 Group C1 C2 C3 1 2 3 4 5 7 8 6...
Страница 128: ...PAG OERLIKON AG CALIBRATION KALIBRIERUNG SERIES 300 SCS B 11 01 9 1 5 8 6 7 3 4 5 2 10 all Groups...
Страница 135: ...PAG OERLIKON AG SERIES 300 S SCS SIP MAIN BOARD 300 7218 010 HAUPTPRINT 300 7218 010 C 01 03...
Страница 150: ...PAG OERLIKON AG SERIES 300 S SCS SIP MAIN BOARD 300 7235 010 HAUPTPRINT 300 7235 010 C 02 06...
Страница 159: ...PAG OERLIKON AG SERIES 300 S SIP ANALOGUE BOARD 300 7219 010 ANALOGPRINT 300 7219 010 C 03 01...
Страница 162: ...PAG OERLIKON AG SERIES 300 SCS ANALOGUE BOARD 300 7238 010 ANALOGPRINT 300 7238 010 C 04 01...
Страница 165: ...PAG OERLIKON AG SERIES 300 S POWER SUPPLY BOARD 300 7215 010 NETZTEILPRINT 300 7215 010 C 05 01...
Страница 168: ...PAG OERLIKON AG POWER SUPPLY BOARD 300 7214 010 NETZTEILPRINT 300 7214 010 SERIES 300 SCS SIP C 06 01...
Страница 194: ...SERIES 300 S SCS SIP PAG OERLIKON AG CALIBRATION INSTRUCTIONS D 04 11 Final control...