<1>. During PID calibration parameter, please don’t operate anything.
<2>. The temp of control has to over room temp about 2ºC for PID calibration parameter.
If the temp of control is lower, please increase it first. Then, customer can decrease the temp
of control after finishing PID calibration parameter.
<3>.After finishing PID calibration parameter,
[AT]
disappear and machine is
normal running automatically. When the temp is not changed during PID calibration
parameter, customer would see alarm [E-15] after3600 sec.
3.9 TEMPERATURE LIMIT AND TIMER
1. During running,
shows CH1 temp.
2. Press
+
2 sec at the same time, enter parameters 2
shows[F01.1] which is
upper limit alarm code of CH1 temp.
3. Press
or
to increase or decrease
the value.
4. Press
to save the current value and enter next item [F02.1] which is lowest limit alarm
code of CH1 temp.
5
.
Press
to save the current value and enter next item [F71] which sets the time of
switch on timely. Customer can set by above operation and back to [F01.1].
6. After completing, press
a while for exiting the setting mode.
7. Parameters2
:
[Temperature 1]
Code
Name
Explanation
Value
Unit
Factory
setting
Range
F01.1
Upper limit of CH1
temp.(Outlet)
Actual temp. > Setting temp. + Temp. upper
limit, PV>SV+[F01.1] alarm and shows [E-08]-
[CH1]
15
0-250
℃
/
℉
F02.1
Lowest limit of CH1
temp. (Outlet)
Actual temp. > Setting temp. – Temp. lowest
limit, PV<SV-[F02.1] alarm and shows [E-07]-
[CH1]
15
0-250
℃
/
℉
F71
Time
Default switching on time.
0000 0~9999
min
13