FirstCall™ Medical Gas Alarm

Operating & Maintenance Manual

Installation

Unpacking Guide

Rough-in box electrical component guide

Rough-in box contains power supply, circuit breaker, 

terminal blocks, input expansions and holes for main 

line power connection and riser connections.

Cover with PLC+HMI

 

PLC+HMI is attached to cover and shipped unattached 

from the rough-in box. Cover should only be attached 

after wall construction is complete. 

Riser Guide (Area & Combo Alarm only)

Risers are type L copper tubing with lead free 

brass  fitting.  Risers  require  gas  specific  DISS  fitting 

with demand valve to be installed prior to installing 

transducers. Risers are included standard with all area 

and combination alarm panels. 

Transducer Guide (Area & Combo Alarm only)

Transducers are unique to each gas. Each transducer 

is labeled appropriate gas and has the corresponding 

DISS hex nut attached to the sensing end. Transducers 

have a solid brass body with ½” NPS threads on cord 

end for easy remote installation.

Medical Gas

Part Number

Oxygen

TDCR-O2

Medical Air

TDCR-AIR

Medical Vacuum

TDCR-VAC

Nitrous Oxide

TDCR-N2O

Nitrogen

TDCR-N2

Carbon Dioxide

TDCR-CO2

WAGD

TDCR-WAGD

Instrument Air

TDCR-INST

Rough-in Install

Rough-in box is shipped with dust cover installed. 

Dust cover will need to be removed to install the rough-

in box and make electrical connections. The dust 

cover should re-installed after the main line electrical 

connection is fully connected to protect the electrical 

components until the wall covering is complete. 

Adjustable  depth  mounting  flanges  (right  &  left) 

should be utilized to align the front edge of the rough-

in box to be flush with the drywall surface. 

Fasten the box to vertical supports (wall studs) so 

that the center of the alarm panel is at appropriate use 

height. 

NOTE: 

Rough-in box should be installed with 

open side facing the outside of the wall and the hinge 

attachment point at the bottom of the panel. 

Reinstall cardboard dust cover.

Содержание AAP10-OAV

Страница 1: ...2AV 15 September 2020 Page 1 of 36 Please read and save these instructions Read carefully before attempting to assemble install operate or maintain the product described Protect yourself and others by...

Страница 2: ......

Страница 3: ...d 5 6 Gas Combo Alarm Panels 16 39 Digital Inputs and 1 4 Gas Combo Alarm Panels 17 39 Digital Inputs and 5 6 Gas Combo Alarm Panels 18 55 Digital Inputs and 1 4 Gas Combo Alarm Panels 19 55 Digital I...

Страница 4: ...onne n ayant pas d exp rience g n rale avec l installation d appareils de cette nature Product Line Overview The Powerex FirstCall medical gas alarm systems monitor the status of medical gases in Cate...

Страница 5: ...Preprogramed standard settings included with all medical gas inputs Alarm set points alarm messages and descriptions of sources are fully customizable Alarm and error history is recorded in Alarm Sum...

Страница 6: ...is labeled appropriate gas and has the corresponding DISS hex nut attached to the sensing end Transducers have a solid brass body with NPS threads on cord end for easy remote installation Medical Gas...

Страница 7: ...to the appropriate terminal block N Hot wire Strip the insulation and connect the hot lead to the appropriate terminal block 120V AC Sensor Installation Area Combo Alarms Only Local Installation Locat...

Страница 8: ...breaker is in the OFF position Attention Pour votre s curit personnelle et pour viter d endommager l alarme assurez vous que le disjoncteur est en position OFF Install the sensor wires into the appro...

Страница 9: ...from power supply into power slot on PLC For panels with 39 digital inputs and or 5 gas inputs Plug I O module expansion adapter cord into appropriate port Place unattached side of panel lanyards over...

Страница 10: ...3 I16 24 I17 25 I18 26 I19 27 I20 28 I21 29 I22 30 15 IN AI3 14 IN AI2 13 COM 1 12 AI1 11 AI0 10 IN 8 9 IN 7 8 IN 6 7 IN 5 6 IN 4 5 IN 3 4 IN 2 3 IN 1 2 IN 0 1 COM 0 Legend A3 A2 A1 A0 YELLOW BLUE BLU...

Страница 11: ...Maintenance Manual Powerex 150 Production Drive Harrison OH 45030 USA P 1 888 769 7979 F 513 367 3125 www powerexinc com IN599102AV 15 September 2020 Page 11 of 36 Area Alarm Panel Wiring Schematic 5...

Страница 12: ...030 USA P 1 888 769 7979 F 513 367 3125 www powerexinc com IN599102AV 15 September 2020 Page 12 of 36 FirstCall Medical Gas Alarm Operating Maintenance Manual Master Alarm Panel Wiring Schematic 23 Di...

Страница 13: ...enance Manual Powerex 150 Production Drive Harrison OH 45030 USA P 1 888 769 7979 F 513 367 3125 www powerexinc com IN599102AV 15 September 2020 Page 13 of 36 Master Alarm Panel Wiring Schematic 39 Di...

Страница 14: ...030 USA P 1 888 769 7979 F 513 367 3125 www powerexinc com IN599102AV 15 September 2020 Page 14 of 36 FirstCall Medical Gas Alarm Operating Maintenance Manual Master Alarm Panel Wiring Schematic 55 Di...

Страница 15: ...19 I18 I17 I16 I15 I14 I13 I12 I11 I10 I9 I8 I7 I6 I5 I4 I3 I2 I1 I0 YELLOW BLUE BLUE A3 A2 A1 A0 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 OUT 6 COM 4 OUT 0 COM 3 BLUE WHITE BLUE W BLUE 14X HORN BLUE W 2X...

Страница 16: ...E 14X BLUE WHITE COM 3 OUT 0 COM 4 OUT 6 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 A0 A1 A2 A3 BLUE BLUE YELLOW I0 I1 I2 I3 I4 I5 I6 I7 I8 I9 I10 I11 I12 I13 I14 I15 I16 I17 I18 I19 I20 I21 I22 FOR BUILDING...

Страница 17: ...1 4 GAS 39 IN BLUE WHITE BLUE WHITE BLUE WHITE BLUE BLUE BLUE BLUE BLUE BLUE WHITE CB N L 24V DC POWER SUPPLY 120V AC 24V DC xx WATTS BLACK RED FOR BUILDING MAIN POWER TIE IN GND LINE N I22 I21 I20 I1...

Страница 18: ...A0 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 OUT 6 COM 4 OUT 0 COM 3 BLUE WHITE BLUE BLUE W BLUE 14X HORN BLUE W 2X BLUE 9X BROWN 4X BLUE 14X BLUE 9X BROWN 4X PLC HMI OUTPUT BAR BLUE WHITE BLUE WHITE 2X 30...

Страница 19: ...I BLUE W 2X BROWN 4X PLC HMI INPUT BAR 2 GRAY BLUE WHITE BLUE W BLUE 9X BLUE 14X 1 2 3 4 5 6 7 12 13 14 15 16 17 18 8 9 10 11 19 20 21 22 1 2 3 4 5 6 7 12 13 14 15 16 17 18 8 9 10 11 19 20 21 22 BLUE...

Страница 20: ...3 12 I3I 11 CM 3 10 9 I2I 8 CM 2 7 6 5 I1I 4 CM 1 3 2 I0I 1 CM 0 I1I I0I BLUE W 2X BLUE WHITE 2X Transd 6 Transd 5 14 13 12 I3I 11 CM 3 10 9 I2I 8 CM 2 7 6 5 I1I 4 CM 1 3 2 I0I 1 CM 0 BLUE 8X GROUP 2...

Страница 21: ...CAPS lock as well as cut copy paste for when you want to use the same information in a different location To maximize the keyboard size to fit the entire screen press the zoom button in the upper lef...

Страница 22: ...dition Use the keyboard to add delete or change as necessary Press the SAVE button at the bottom to save changes Press the SETTINGS MENU button to return to the settings screen or the HOME button to r...

Страница 23: ...nd Combo alarms In the gas settings page you are able to change the medical gas type add subtract a gas designate the location monitored for each gas and manually change the high low alarm settings NO...

Страница 24: ...t one Press SAVE SOURCE underneath to add the new source badge to the list of Source Types The table to the right will then be populated with selectable boxes and a heading with the selected source ga...

Страница 25: ...ring a normal alarm event This allows a technician to perform maintenance without setting off the loud alarm Maintenance mode will automatically toggle back to normal after 15 minutes A timer is displ...

Страница 26: ...rex Technical Service for instruction Test Set High Low Alarm Points to 20 Current Pressure This unique feature easily allows for compliance with NFPA 99 2018 edition 5 1 9 4 2 stating that Area Alarm...

Страница 27: ...is the wrong gas type 3 Low gas pressure less than 20 below nominal Operation TEST Feature To test the alarm horn press the TEST button in the center of the lower bar A caution window will pop up warn...

Страница 28: ...of alarm conditions The specific alarm will be highlighted in red Silencing the Alarm In the event of an alarm pressing the silence button will silence the horn indefinitely The scrolling marquee at t...

Страница 29: ...use a microfiber cloth or soft lint free cloth when cleaning any smudges off of the HMI Do not use a paper towel or tissue paper wet or dry Do not apply an excessive amount of pressure to the screen...

Страница 30: ...arbon Dioxide WAGD Instrument Air TDCR O2 TDCR AIR TDCR VAC TDCR N2O TDCR N2 TDCR CO2 TDCR WAGD TDCR INST Terminal Blocks PE000652AV Snap in Plugs SL057170AV Micro SD Card PE000467AV Horn PE001011AV R...

Страница 31: ...Faulty power supply Replace power supply Wiring harness disconnected Make sure the wiring harness from the power supply and terminal blocks are properly installed in the back of the PLC Alarm panel f...

Страница 32: ...ts Input capable of 24 VDC 120 VAC DC output Dry Contacts capable of 30 VDC or 120 VAC max output Input fuse 2 Amp Power consumption 30 W Maximum Pressure measurement accuracy All positive pressure tr...

Страница 33: ...larm G7HiAlarm G8HiAlarm Appendix A BACnet Object Names BACnet Object Name Object Name Description G1HiAlarm Gas Badge High Alarm Status A displayed value of 0 Normal 1 Alarm G2HiAlarm G3HiAlarm G4HiA...

Страница 34: ...rex 150 Production Drive Harrison OH 45030 USA P 1 888 769 7979 F 513 367 3125 www powerexinc com IN599102AV 15 September 2020 Page 34 of 36 FirstCall Medical Gas Alarm Operating Maintenance Manual No...

Страница 35: ...CONNECTION WITH THE SALE OR USE OF SUCH PRODUCTS TO THE EXTENT PERMITTED BY LAW ANY AND ALL IMPLIED WARRANTIES ARE EXCLUDED All warranty claims must be made in writing and delivered to Powerex in acc...

Страница 36: ...indirect special or consequential damages or loss of production of loss of profits resulting from any cause whatsoever including but not limited to any delay act error or omission of Powerex Supplier...

Отзывы: