FirstCall™ Medical Gas Alarm
Operating & Maintenance Manual
Powerex • 150 Production Drive • Harrison, OH 45030 • USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
IN599102AV • 15 September 2020
Page 15 of 36
Combo Alarm Panel Wiring Schematic
23 Digital Inputs and 1-4 Gas Combo Alarm Panels – AP000C234AJ
TRANS.
1
RED
BLA
CK
TRANS.
2
TRANS.
3
TRANS.
4
RED
BLA
CK
RED
BLA
CK
RED
BLA
CK
RED (4X)
Note:Transducers and transducer wiring is field installed
BLACK (4X)
BLUE-WHITE
BLUE
BLUE
BLUE
BLUE
WHITE
CB
N
L
24V
DC
-
+
POWER
SUPPLY
120V AC
24V DC
xx WATTS
BLACK
RED
FOR BUILDING MAIN POWER TIE IN
GND LINE N
I22 I21 I20 I19 I18 I17 I16 I15 I14 I13 I12 I11 I10 I9 I8 I7 I6 I5
I4 I3 I2 I
1
I
0
YELLOW
BLUE
BLUE
A3
A2
A1
A0
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
OUT
6
COM
4
OUT
0
COM
3
BLU
E-
W
HI
TE
BLUE-W, BLUE (14X)
-
+
HORN
BLUE-W (2X), BLUE (9X), BROWN (4X)
BLUE (14X)
BLUE (9X)
BROWN (4X)
PLC/HMI
OUTPUT BAR
BLUE-WHITE
BLUE (14X)
BLUE (9X)
BLUE-W
BLUE-WHITE
PLC/HMI
INPUT BAR 2 (GRAY)
BROWN (4X)
BLUE-W (2X)
PLC/HMI
BLUE
GREEN
BLUE-WHITE
0V
+V
PLC/HMI
INPUT BAR 1 (BLACK)
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
AI0
AI1
COM
0
IN
0
IN
1
IN
2
IN
3
IN
4
IN
5
IN
6
IN
7
IN
8
COM
1
IN
AI2
IN
AI3
30
29
28
27
26
25
24
23
22
21
20
19
18
17
16
I22
I21
I20
I19
I18
I17
I16
I15
I14
I13
I12
I11
I10
I9
COM2
All Models of Powerex FirstCall alarm panels must be powered by the life safety branch or
critical branch of the essential electrical system
as required in NFPA 99