FirstCall™ Medical Gas Alarm
Operating & Maintenance Manual
4. High – gas pressure greater than 20% above
nominal.
Master Alarm – Any Signal From the Source
Equipment
In the event of an alarm on a Master Alarm panel, the
source gas that is experiencing the alarm will change
from the green “NORMAL” box to a flashing red “INPUT”
box.
To see the specific alarm, press the gas badge that’s in
alarm to view the list of alarm conditions. The specific
alarm will be highlighted in red.
Silencing the Alarm
In the event of an alarm, pressing the silence button
will silence the horn indefinitely. The scrolling marquee
at the buttom will remain.
Scrolling Marquee
The scrolling marquee displays all active alarm
conditions.
Zzz button – “snooze” button, only minimizes the
scrolling marquee.
Scrolling marquee will remain until:
• Alarm condition is remedied or acknowledged
– acknowledge by pressing the Alarm Summary
button (second from the left) to get to the Alarm
Summary Screen.
•
Pressing the flashing bell icon at the very left will
temporarily hide the scrolling marquee, but unless
the alarm condition is remedied or acknowleged,
it will reappear after a short period of time.