GUIDANCE AND MANUFACTURER’S DECLARATION –
ELECTROMAGNETIC IMMUNITY
PowerDot PD-01 device is intended for use in the electromagnetic
environment specified below. The customer or the user of the PowerDot
PD-01 device should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment –
Guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
±
6kV contact
±
8kV air
±
6kV
contact
±
8kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±
2kV for power
supply lines
±
1kV for input/
output lines
Not
Applicable
PowerDot PD-01 does
not connect to AC
source directly.
Surge IEC
61000-4-5
±
1kV Line to Line
(or Neutral)
±2kV Line(s) to
earth
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input units
IEC 61000-4-11a
<5% UT (>95%
dip in UT)
for 0.5
Cycle
40% UT (60%
dip in UT) for 5
Cycles
70% UT (30%
dip in UT) for 25