Classification
1. Class 1 equipment
2. Type B applied part;
3. IPX0;
4. Not category AP / APG equipment;
5. Mode of operation: Non-continuous operation: 9 min on / 3 hours off
Environmental Requirements
OPERATING CONDITIONS
Temperature: 10ºC to 40ºC
Humidity: 10% to 95% RH
Pressure altitude: Normal atmospheric
conditions/ 860-1060hPa
STORAGE AND SHIPPING CONDITIONS
Temperature: 10ºC to 40ºC
Humidity: 10% to 95% RH
Pressure altitude: Normal atmospheric
conditions/ 860-1060hPa
WARNING
•
The device is not suitable for use in the presence of flammable anesthetic mixtures with air or with
oxygen or nitrous oxide.
•
The device requires no calibration.
•
The device is not repairable and contains no user serviceable parts.
•
The user must check that the equipment functions safely and see that it is in proper working
condition before being used.
•
Permissible environmental conditions of use including conditions for transport and storage
•
The manufacturer does not require such preventive inspections by other persons.
•
The plug is used to disconnect from the main supply. Do not position the machine so that it is
difficult to disconnect the main plug.
•
Warning: To avoid the risk of electric shock, this equipment must only be connected to a supply
main with protective earth grounding.
•
This appliance is not intended for use by persons (including children) with reduced physical,
sensory or mental capabilities, or lack of experience and knowledge, unless they have been given
supervision or instruction concerning use of the appliance by a person responsible for their safety.
•
Children should be supervised to ensure that they do not play with the appliance.
•
If the supply cord is damaged, it must be replaced by the manufacturer, its service agent or
similarly qualified persons in order to avoid a hazard.
Patient
The machine is intended for use by a diverse patient population including:
• Adult men and woman who are medically cleared and able to undertake physical exercises such
as weight baring and ground reaction force type exercise programs. In general this applies to
healthy grownups (16+ years old).
• All others users should be cleared by their physician and only use the Power Plate under medical
supervision.
Performance
The machine is intended to be used in the following environments:
• Commercial Medical Facilities
• Health Clinics
• Hospitals
• Rehabilitation Centers