Guidance and manufacture’s declaration – electromagnetic immunity –
for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The Pro5M is intended for use in the electromagnetic environment specified below. The customer or the user of
Pro5M should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic
±6 kV contact
±6 kV contact
Floors should be wood, concrete or
discharge (ESD)
±8 kV air
±8 kV air
ceramic tile. If floor are covered with
IEC 61000-4-2
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
±2 kV for power
±2kV for power
Mains power quality should be that of
transient/burst
supply lines
supply lines
a typical commercial or hospital IEC
61000-4-4
environment.
Surge
± 1 kV line(s) t
±1 kV differential
Mains power quality should be that of
IEC 61000-4-5
o line(s)
mode
a typical commercial or hospital
environment.
Voltage dips, short
<5% U
T
<5% U
T
Mains power quality should be that
interruptions and
(>95% dip in U
T
)
(>95% dip in U
T
)
of a typical commercial or hospital
voltage variations
for 0.5 cycle
for 0.5 cycle
environment. If the user of the Pro5M
on power supply
requires continued operation during
input lines
40% U
T
40% U
T
power mains interruptions, it is IEC
61000-4-11
(60% dip in U
T
)
(60% dip in U
T
)
recommended that the Pro5M be powered
for 5 cycles
for 5 cycles
from an uninterruptible power supply or a
battery.
70% U
T
70% U
T
(30% dip in U
T
)
(30% dip in U
T
)
for 25 cycles
for 25 cycles
<5% U
T
<5% U
T
(>95% dip in U
T
)
(>95% dip in U
T
)
for 5 sec
for 5 sec
Power frequency
3A/m
3A/m
Power frequency magnetic fields should
(50Hz) magnetic
be at levels characteristic of a typical
61000-4-8
location in a typical commercial or
hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.