Guidance and manufacture’s declaration – electromagnetic immunity –

for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Guidance and manufacture’s declaration – electromagnetic immunity

The Pro5M is intended for use in the electromagnetic environment specified below. The customer or the user of 
Pro5M should assure that it is used in such an environment.

Immunity test 

IEC 60601 test 

Compliance  

Electromagnetic environment 

 

 

level 

level  - 

guidance 

   

   

 

 

Portable and mobile RF communications  

   

 

 

equipment should be used no closer to  

   

 

 

any part of the Pro5M, including cables, than  

   

 

 

the  recommended separation distance  

   

 

 

calculated from the equation applicable to the  

   

 

 

frequency of the transmitter.

   

 

 

Recommended separation distance

Conducted RF 

3 Vrms 

3 Vrms

IEC 61000-4-6 

150 kHz to 80 MHz

Radiated RF 

3 V/m 

3 Vrms 

 

            80 MHz to 800 MHz

IEC 61000-4-3 

80 MHz to 2.5 GHz 

 

 

            800 MHz to 2.5 GHz

  

 

 Where 

P is the maximum output power rating  

   

 

 

of the transmitter in watts (W) according  

   

 

 

to the transmitter manufacturer and d is the  

   

 

 

recommended separation distance in metres (m).

   

 

 

Field strengths from fixed RF transmitters, as  

   

 

 

determined by an electromagnetic site survey,

a

  

   

 

 

should be less than the compliance level in each  

   

 

 

frequency range.

b

   

 

 

Interference may occur in the vicinity of  

   

 

 

equipment marked with the following symbol: 

 
 
 

NOTE 1   At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2   These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 
and reflection from structures, objects and people.

a

   Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 
survey should be considered. If the measured field strength in the location in which the Pro5M is used exceeds 
the applicable RF compliance level above, the  Pro5M should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as reorienting or relocating the Pro5M.

 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

P

d

167

.

1

=

P

d

167

.

1

=

P

d

167

.

1

=

14

Содержание Pro5M

Страница 1: ...English Instructions for Use Pro5M Model Medical Device...

Страница 2: ...o5HP My7M my7 my5 my3 Personal MOVE Pro4PlusM and proMOTION are registered trademarks and or trademarks of Peformance Health Systems and or their affiliates All other trademarks are the property of th...

Страница 3: ...int dizzy nauseous short of breath or in pain stop exercising immedia tely and consult with your physician Body posture muscle stiffness and muscle tension are important contributing factors in your t...

Страница 4: ...ectively make up the circulatory system Pulmonary circulation is the portion of the cardiovascular system which transports oxygen depleted blood away from the heart to the lungs and returns oxygenated...

Страница 5: ...variation and the shape of the bones and cartilage that make up the joint Pathology Studies Claim Power Plate training can be a beneficial tool intervention to improve general well being fitness and d...

Страница 6: ...ing on and off the machine Never reach into or under the machine or tip the machine on its side while it is in operation Use the machine only for the purposes described in this user manual and only wi...

Страница 7: ...ssible injury from the machine s vibrations Power Plate recommends that the user not lean back on his her heels or lock his her joints or straighten his her legs when standing upright on the machine I...

Страница 8: ...are used in such an environment Floors should be wood concrete or ceramic tile If the floor is covered with synthetic material the relative humidity in the room should be at least 30 Mains power qual...

Страница 9: ...machine so that it is difficult to disconnect the main plug Warning To avoid the risk of electric shock this equipment must only be connected to a supply main with protective earth grounding This appl...

Страница 10: ...Operation Power Plate pro5 face plate 10...

Страница 11: ...in your training The Power Plate machine is very easy to operate as you can tell from a brief review of the primary buttons on the face plate 1st Main Controls Start Repeat 2nd Frequency 25 50Hz 1Hz i...

Страница 12: ...unction Therefore its RF emissions are very low and are not likely to cause any interfe rence in nearby electronic equipment RF emission CISPR 11 Class B Class A with IEC61000 3 2 Complies with IEC610...

Страница 13: ...nment Surge 1 kV line s t 1 kV differential Mains power quality should be that of IEC 61000 4 5 o line s mode a typical commercial or hospital environment Voltage dips short 5 UT 5 UT Mains power qual...

Страница 14: ...eld strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of...

Страница 15: ...42 EEC Date of manufacture Manufacturer Type B applied part DISPOSAL Do not dispose this product as unsorted municipal waste Collection of such waste separately for special treatment is necessary CB C...

Страница 16: ...ises can be added When trying to accomplish weight loss cardiovascular or endurance improvements rest time can be reduced exercise time can be extended and the number of exercises can be increased to...

Страница 17: ...cademies For more information please visit our website www powerplate com PersonalAdaptation Training on the Power Plate machine is like any other type of training start with light and short training...

Страница 18: ...rtz 30 45 or 60 seconds Up to 9 minutes TimeSelections VibrationEnergyOutput Low or High DimensionsWxDxH 96 x 114 x 153 cm 38 x 45 x 60 SurfacePlateDimensionsWxD 96 x 114 cm 38 x 45 Weight 185 kilogra...

Страница 19: ...tions are proper please check the circuit breaker or fuse for the electrical outlet that is supplying power to the machine in order to ensure that the outlet is receiving electrical power If the conso...

Страница 20: ...y be obtained by contacting Power Plate through the Contact Us page online or by writing to Power Plate Performance Health Systems UK Ltd Business Design Centre Unit 115 52 Upper Street Islington Lond...

Отзывы: