7. Self Inspection and Maintenance
7.1 Before Calling a Service Person
Warning messages will be displayed on the monitor if some problems occur. It might be
operation errors or problems of the machine.
In this case, refer the following instructions.
If the function is still not salvaged or recovered, disconnect the power supply and consult the
dealer.
(1) Message When Power On.
Message Cause
Remedy
REF/KER CCD CHECK...
Internal error
Turn OFF the power switch and turn
on again after 10 seconds.
If the message appears again,
consult the dealer.
REF MOTOR ORIGIN... FAILED
INVALID SETUP DATA - REF
Internal Setup Data for
Refractometry is invalid.
Consult the dealer.
INVALID SETUP DATA - KER
Internal Setup Data for Keratometry
is invalid.
Consult the dealer.
(2) Message On Measuring
Message Cause
Remedy
ERROR
Alignment is improper
Measure after aligning the pupil and the
Alignment Mark properly.
Eyelid or eyelashes are covering the pupil.
Instruct the examinee to open his or her eyes
wide, or lift up the eyelid lightly with your fingers
and measure again
When the pupil is smaller than the Outer
Alignment Mark.
The minimum pupil diameter that can be
measured is 2.0 mm. Although it is possible to
measure in the bright place, don’t expose
examinee’s eyes to the direct sunlight or too
bright indoor lights to prevent the contraction of
the pupil.
When the examinee has some illness like
cataract.
Observe the eye in RET Mode. If cataract is not
severe, measurement can be performed in the
IOL mode.
Examinee has IOL(intra-ocular lens)
implanted.
Measure in the IOL mode.
When the Mire Image is odd shaped
because of tears.
Instruct the examinee to open and close his or
her eyes several times and measure again.