ENGLISH
81
Additional NTC probe and RJ45 cabel (client’s probe, accessory)
F
• Remove the protective cap (E)
• Twist to open the bushing of the push through plug (F)
• Push sensor (G) through the push through hole into the
box
• Plug in the cables into the switch strip (C)
• Twist to close the push through plug (F)
Note!
Customer sensor must be calibrated according to the box
temperatures. Sensor calibration depends on its location.
DIN 58345 accessory package for MED SAVE control unit
There is a potential-free connection point (A) at the back of the cabinet for alarms. DIN 58345 equipment package
includes a battery backup for MED SAVE controller power failure and temperature registering during 12 hours, lockable
battery back-up user switch (B), and antifreeze function.
BB cables have battery back-up by default.