background image

 
 

 

WWW.PME.COM 

TECHNICAL SUPPORT: [email protected] | (760) 727-0300 

PG. 24 

THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED. 

 

 

CHAPTER 3: MINIWIPER 

3.1 Overview 

The miniWIPER is intended to provide a gentle wiping of the Cyclops-7oxygensensitive membrane 
at user-programmable intervals. Customers will be required to open the wiper to set the wiper 
interval or to exchange the batteries. This chapter describes the wiper internal features. 

3.2 Opening and Closing the Wiper 

The wiper circuitry is contained in a waterproof housing that must be opened. The housing is opened 
by unscrewing the white pressure housing from the black end cap. Turn the pressure housing 
counter clockwise relative to black end cap. Close by reversing this procedure after being sure that 
the o-ring is free from debris. Lube o-ring occasionally with grease intended for buna-N o-ring 
material. 

Please attempt to handle the miniWIPER only by the aluminum chassis, without touching the circuit 
card. 

When closing the miniWIPER just screw the white cylinder onto the black end cap until the cylinder 
just touches the end cap. Do not tighten! miniWIPER tends to get a little tighter during deployment. 

3.3 Electrical Connections and Controls  

Removal of the cover reveals the wiper connections and controls, shown below. 

 

The 

LED Light

 is a red or green LED. This is used to indicate different features described elsewhere 

in this manual. 

 

 

Содержание Cyclops-7

Страница 1: ...PORT INFO PME COM 760 727 0300 PG 1 THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL 2021 PRECISION MEASUREMENT ENGINEERING INC ALL RIGHTS RESERVED 1483 POINSETTIA AVE STE 101 VISTA CA 92081 USA Cyclops...

Страница 2: ...5 Bursting Hazard 5 CHAPTER 1 INSTALLATION 6 1 1 Overview 6 1 2 Installation Instructions 7 1 3 Wipe Considerations 9 CHAPTER 2 SOFTWARE 16 2 1 Overview and Software Installation 16 2 2 miniWIPERCont...

Страница 3: ...ogger Logger only 1 year For valid warranty claims made and covered defects existing during the applicable warranty period PME will at PME s option repair replace with the same or then most similar pr...

Страница 4: ...e for any indirect incidental special exemplary or consequential damages including but not limited to lost profits loss of data loss of use business interruption loss of good will or cost of procuring...

Страница 5: ...the miniWIPER and meet the enclosed batteries the batteries may generate gas causing the internal pressure to increase This gas will likely exit via the same location where the water entered but not n...

Страница 6: ...e wiper is set to wipe every 6 hours Wipe interval can be selected by the customer with intervals as short as 1 hour or as long as 24 hours Wipe intervals outside this range are available Please conta...

Страница 7: ...ds the Cyclops 7 sensor in place must remain installed Install the included wiper sensor guard to replace the sensor shade The bracket will grip the sensor retainer and the wiper sensor guard Construc...

Страница 8: ...Y BY HAND CAN DAMAGE miniWIPER We also suggest aligning the miniWIPER by using the included magnet By waving the magnet over the miniWIPER it will automatically begin a wipe Tighten the 6 bolts while...

Страница 9: ...t be less than or equal to 24 hours Intervals outside this range will be rejected by the miniWIPERControl program Contact PME for other recording intervals Please refer to Chapter 2 for instructions o...

Страница 10: ...E The miniWIPER consumes battery energy mostly from operating the wiper motor but also slightly from simply keeping track of time writing files sleeping and other activities The amount of energy consu...

Страница 11: ...ince miniWIPER draws 50 to 100 mA depending on various factors but the general shape of the voltage vs time gives an estimate of life remaining This plot is taken from the manufacturer s specification...

Страница 12: ...0 PG 12 THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL 2021 PRECISION MEASUREMENT ENGINEERING INC ALL RIGHTS RESERVED Duracell alkaline AA at room temperature 7533 single scrub wipes Energizer L91 Lith...

Страница 13: ...he o ring sealing the drive shaft The brush will be replaced by the customer by replacing the brush wheel The o ring can be replaced by the customer We do not know the replacement interval at this tim...

Страница 14: ...asurements This indication appears in response to switching the Logger Control Switch to Halt Continuously Green Normal Indicates the Cyclops 7 Logger is connected to a HOST computer via the USB conne...

Страница 15: ...pter with a micro USB connector on one end that connects to your Android device and a standard A type USB plug on the other The cable supplied with the miniWIPER connects to this end 3 A File Explorer...

Страница 16: ...miniWIPERConcatenate jar gathers all daily files into one CAT txt file Manual pdf is this manual These files are located on the root directory of the miniWIPER This appears as a Local Disk when miniW...

Страница 17: ...e than a week has passed since time was last set the miniWIPER clock will be set and a check mark icon will appear If the HOSTcomputer is not connected to the internet no time services will occur The...

Страница 18: ...ogram will suggest the folder that is located on the miniWIPER s SD card You may accept this by clicking Plot or you may click Select DATA Folder to browse to your HOSTcomputer s hard drive If the num...

Страница 19: ...WWW PME COM TECHNICAL SUPPORT INFO PME COM 760 727 0300 PG 19 THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL 2021 PRECISION MEASUREMENT ENGINEERING INC ALL RIGHTS RESERVED...

Страница 20: ...WWW PME COM TECHNICAL SUPPORT INFO PME COM 760 727 0300 PG 20 THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL 2021 PRECISION MEASUREMENT ENGINEERING INC ALL RIGHTS RESERVED...

Страница 21: ...already processed DATA Folder is selected then the program simply reads the miniWIPER s measurement files again End the miniWIPERPlot program by closing the window In general the recorded information...

Страница 22: ...IPE Please review the Low Drain Performance figure in section 1 3 The lithium batteries can provide 0 050 Amp service for roughly 50 hours at 1 4 Volts miniWIPER uses two batteries so the miniWIPER ba...

Страница 23: ...PER If the miniWIPERPlot program is run directly from the miniWIPER then the program will suggest the folder located on the miniWIPER You may accept this by clicking Concatenate or you may click Selec...

Страница 24: ...screwing the white pressure housing from the black end cap Turn the pressure housing counter clockwise relative to black end cap Close by reversing this procedure after being sure that the o ring is f...

Страница 25: ...Energizer L91 AA size batteries You may also use AA size alkaline batteries such as Duracell Coppertop http data energizer com PDFs l91 pdf http ww2 duracell com media en US pdf gtcl Product_Data_She...

Страница 26: ...tend vertically downward to the next sensor wiper or to an anchor The rope from the opposite end will extend vertically upward to the next sensor wiper or to a surface float See the sensor manual for...

Отзывы: