2.2 miniWIPERControl 

Begin program operation by clicking on 
miniWIPERControl.jar. Software presents the 
screen to the left (shown after miniWIPER is 
connected):   

The miniWIPER must be connected to USB at this 
time. When correctly connected the miniWIPER 
LED will display a constant green light. 

Click the Connect Button. The software will contact 
the wiper. If the connection is successful, the button 
will turn green and display “Connected” as seen on 
the left. The Serial Number and other parameters 
will be filled in from information taken from the 
miniWIPER. 

If the HOSTcomputer is connected to the internet, 
the current difference between an internet time 
server’s time and the miniWIPER internal clock will 

be displayed. And, if more than a week has passed since time was last set, the miniWIPER clock will 
be set and a check mark icon will appear. If the HOSTcomputer is not connected to the internet no 
time services will occur. 

The current miniWIPER wipe interval will be displayed next to the Set Wipe Interval button. If this 
interval is acceptable the interval need not be set. 

To set the interval, enter an interval not less than 1 hour and not greater than 24 hours. Click the Set 
Sample Interval button. Shorter and longer intervals are available. Contact PME. 

Click the test wiper button to operate the wiper motor briefly. The wiper will make one complete wipe 
over a period of about 30 seconds. (A complete wipe is rotation in one direction only. Note that 
miniWIPER occasionally does a calibration wipe in addition to a scrub wipe. A calibration wipe is a 
wipe to both rotational endpoints without scrubbing.) 

End miniWIPER Control by closing the window. Unplug miniWIPER USB connection. 

Upon disconnection of the USB cable the miniWIPER will begin wiping or remain halted as indicated 
by the position of the Wiper Control Switch. 

 

 

Содержание Cyclops-7

Страница 1: ...PORT INFO PME COM 760 727 0300 PG 1 THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL 2021 PRECISION MEASUREMENT ENGINEERING INC ALL RIGHTS RESERVED 1483 POINSETTIA AVE STE 101 VISTA CA 92081 USA Cyclops...

Страница 2: ...5 Bursting Hazard 5 CHAPTER 1 INSTALLATION 6 1 1 Overview 6 1 2 Installation Instructions 7 1 3 Wipe Considerations 9 CHAPTER 2 SOFTWARE 16 2 1 Overview and Software Installation 16 2 2 miniWIPERCont...

Страница 3: ...ogger Logger only 1 year For valid warranty claims made and covered defects existing during the applicable warranty period PME will at PME s option repair replace with the same or then most similar pr...

Страница 4: ...e for any indirect incidental special exemplary or consequential damages including but not limited to lost profits loss of data loss of use business interruption loss of good will or cost of procuring...

Страница 5: ...the miniWIPER and meet the enclosed batteries the batteries may generate gas causing the internal pressure to increase This gas will likely exit via the same location where the water entered but not n...

Страница 6: ...e wiper is set to wipe every 6 hours Wipe interval can be selected by the customer with intervals as short as 1 hour or as long as 24 hours Wipe intervals outside this range are available Please conta...

Страница 7: ...ds the Cyclops 7 sensor in place must remain installed Install the included wiper sensor guard to replace the sensor shade The bracket will grip the sensor retainer and the wiper sensor guard Construc...

Страница 8: ...Y BY HAND CAN DAMAGE miniWIPER We also suggest aligning the miniWIPER by using the included magnet By waving the magnet over the miniWIPER it will automatically begin a wipe Tighten the 6 bolts while...

Страница 9: ...t be less than or equal to 24 hours Intervals outside this range will be rejected by the miniWIPERControl program Contact PME for other recording intervals Please refer to Chapter 2 for instructions o...

Страница 10: ...E The miniWIPER consumes battery energy mostly from operating the wiper motor but also slightly from simply keeping track of time writing files sleeping and other activities The amount of energy consu...

Страница 11: ...ince miniWIPER draws 50 to 100 mA depending on various factors but the general shape of the voltage vs time gives an estimate of life remaining This plot is taken from the manufacturer s specification...

Страница 12: ...0 PG 12 THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL 2021 PRECISION MEASUREMENT ENGINEERING INC ALL RIGHTS RESERVED Duracell alkaline AA at room temperature 7533 single scrub wipes Energizer L91 Lith...

Страница 13: ...he o ring sealing the drive shaft The brush will be replaced by the customer by replacing the brush wheel The o ring can be replaced by the customer We do not know the replacement interval at this tim...

Страница 14: ...asurements This indication appears in response to switching the Logger Control Switch to Halt Continuously Green Normal Indicates the Cyclops 7 Logger is connected to a HOST computer via the USB conne...

Страница 15: ...pter with a micro USB connector on one end that connects to your Android device and a standard A type USB plug on the other The cable supplied with the miniWIPER connects to this end 3 A File Explorer...

Страница 16: ...miniWIPERConcatenate jar gathers all daily files into one CAT txt file Manual pdf is this manual These files are located on the root directory of the miniWIPER This appears as a Local Disk when miniW...

Страница 17: ...e than a week has passed since time was last set the miniWIPER clock will be set and a check mark icon will appear If the HOSTcomputer is not connected to the internet no time services will occur The...

Страница 18: ...ogram will suggest the folder that is located on the miniWIPER s SD card You may accept this by clicking Plot or you may click Select DATA Folder to browse to your HOSTcomputer s hard drive If the num...

Страница 19: ...WWW PME COM TECHNICAL SUPPORT INFO PME COM 760 727 0300 PG 19 THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL 2021 PRECISION MEASUREMENT ENGINEERING INC ALL RIGHTS RESERVED...

Страница 20: ...WWW PME COM TECHNICAL SUPPORT INFO PME COM 760 727 0300 PG 20 THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL 2021 PRECISION MEASUREMENT ENGINEERING INC ALL RIGHTS RESERVED...

Страница 21: ...already processed DATA Folder is selected then the program simply reads the miniWIPER s measurement files again End the miniWIPERPlot program by closing the window In general the recorded information...

Страница 22: ...IPE Please review the Low Drain Performance figure in section 1 3 The lithium batteries can provide 0 050 Amp service for roughly 50 hours at 1 4 Volts miniWIPER uses two batteries so the miniWIPER ba...

Страница 23: ...PER If the miniWIPERPlot program is run directly from the miniWIPER then the program will suggest the folder located on the miniWIPER You may accept this by clicking Concatenate or you may click Selec...

Страница 24: ...screwing the white pressure housing from the black end cap Turn the pressure housing counter clockwise relative to black end cap Close by reversing this procedure after being sure that the o ring is f...

Страница 25: ...Energizer L91 AA size batteries You may also use AA size alkaline batteries such as Duracell Coppertop http data energizer com PDFs l91 pdf http ww2 duracell com media en US pdf gtcl Product_Data_She...

Страница 26: ...tend vertically downward to the next sensor wiper or to an anchor The rope from the opposite end will extend vertically upward to the next sensor wiper or to a surface float See the sensor manual for...