ERROR MESSAGES
30 Planmeca Intra X-ray unit
Technical Manual
E.71
FLASH memory check-sum error (tube head CPU)
Tube head CPU internal error.
Replace the tube head PCB.
E.81
EEPROM memory defective (tube head CPU)
Tube head CPU internal error.
Replace the tube head PCB.
E.83
Config register error (tube head CPU)
Tube head CPU internal error.
Replace the tube head PCB.