SERVICE MODE
2.9
Disabling the exposure key
Enter the service mode according to the instruction given in section 2.2 “How to enter/exit
the service mode” on page 12.
Press the parameter adjustment up key until the parameter number 18 appears on the kV
display.
The number indicating the mode of exposure key operation is shown on the time display. 0
= exposure key normal operation, 1 = exposure key operation disabled. The factory
default is 0.
Press the Select key until the number starts to blink, and the value can now be changed
with the parameter adjustment keys.
Accept the new setting by pressing the Select key.
Press the Mode key briefly to exit the service mode.
2.10
Automatic identification of the cascade card
Enter the service mode according to the instruction given in section 2.2 “How to enter/exit
the service mode” on page 12.
Press the parameter adjustment up key until the parameter number 19 appears on the kV
display.
The value indicating the type of the cascade card is shown on the time display. Value <
108 equals an old card and value
≥
108 equals a new card. The type of the cascade card
affects the adjustment functionality of the preheat and mA.
Parameter adjustment up -key
18
0
Parameter adjustment keys
Select key
Parameter adjustment up -key
19
108