Planmeca Intra X-ray unit 15
Technical Manual
SERVICE MODE
be changed automatically. If the exposure time is manually altered, the unit will return into
manual mode. At any time, the preprogrammed setting for the kV value, mA value and
exposure time for any tooth position is called by selecting the preprogrammed setting and
pressing the child mode selection key twice.
Press and hold down the Select key until the number starts to blink, and the number can
now be changed with the parameter adjustment keys.
Accept the new setting by pressing the Select key.
Press the Mode key briefly to exit the service mode.
2.7
Dimming the displays
Enter the service mode according to the instruction given in section 2.2 “How to enter/exit
the service mode” on page 12.
Press the parameter adjustment up key until the parameter number 16 appears on the kV
display.
The number indicating the brightness of the displays (from 1 to 5, factory default is 5) is
shown on the time display.
Press the Select key until the number starts to blink, and the value can now be changed
with the parameter adjustment keys.
Accept the new setting by pressing the Select key.
Press the Mode key briefly to exit the service mode.
Parameter adjustment keys
Select key
Parameter adjustment up -key
16
5
Parameter adjustment keys
Select key
Содержание intra
Страница 1: ...technical manual 10006491_1 EN ...
Страница 2: ......
Страница 53: ...Planmeca Intra X ray unit 49 Technical Manual DIAGRAMS 9 DIAGRAMS ...
Страница 54: ......
Страница 56: ......
Страница 57: ......