3.c. Attach the Dürr contact to the suction element end of the suction
tube (3).
3.d. Attach the suction valve to the suction tube (5) and attach the
suction tube to the corresponding tube coming from the floor (7).
Secure the tubes into place with securing rings (4 and 6).
4. Install the assistant's syringe.
NOTE
The assistant's syringe is an optional feature.
4.a. Route the assistant syringe cable (1) and the assistant syringe
water and air tubes (2) from the opening for the suction element
through the lifting mechanism to the junction box.
In the lifting mechanism, the suction tube is placed on the upper
cable duct (2).
The assistant syringe quick-connector (4) will later be attached to
the suction element with two attachment screws (5), see step 5d.
4.b. Connect the assistant syringe cable to the Main control PCB (6).
4.c. Connect the assistant syringe air tube (2) to the air filter/regulator.
4.d. Connect the assistant syringe water tube (2) to the assistant
syringe valve.
12 Installing suction element
62 Planmeca Compact i3
Installation manual