Ethernet cable
The Ethernet cable is needed for dental unit / Planmeca Romexis server
connection.
USB cable
The USB cable is needed for USB intraoral camera connection. The
repeater cable is routed from the dental unit to the PC. If the distance is over
5 m, another repeater cable must be added.
Other cables
2.4.10 Network connection
An Ethernet network is used to connect to the Romexis Clinic Management
for equipment diagnostics and remote user assistance.
For more information on network connection requirements, see section
"Network architecture requirements" on page 8.
2.5 Network architecture requirements
The dental unit requires a network connection for the purposes of connecting
with Planmeca Romexis.
The user organisation must take care of protecting the network by using an
up-to-date virus and malware protection software and a firewall. The network
must be private and access to the network must be strictly restricted. The
network must not be connected to the Internet.
The recommended connection speed is 100 Mb/s.
For instructions on how to enable remote connection (PMUAPI) and
configure the unit’s IP address, see technical manual section
Network
settings.
CAUTION
Never connect the unit to the Internet. Connecting the unit to Internet will
result in previously unidentified risks to patients, operators or third parties.
Connect the unit to private network only.
CAUTION
Never turn on unit power while having your fingers in patient chair
mechanics. Remote user at Romexis Clinic Management may start chair
automatic movements via Ethernet connection.
2 Pre-installation requirements
8 Planmeca Compact i3
Installation manual