Headrest related error messages
ERROR
CODE
EXPLANATION
E80.6
Motorised headrest motor B potentiometer or cable is faulty or
disconnected.
E80.7
Motorised headrest length potentiometer or cable is faulty or
disconnected.
E81
Motorised headrest communication error (may be too old or
wrong software or loose cables or short circuit).
E81.1
Motorised headrest communication error.
E82
Headrest software is incompatible with MCB software.
Headrest or MCB software should be upgraded.
E82.1
Headrest software is incompatible with MCB software.
Headrest or MCB software should be upgraded.
E82.2
Incompatible headrest software.
31.3.13 ActiveAqua related error messages
ActiveAqua related error messages
ERROR
CODE
EXPLANATION
E83.1
ActiveAqua and main PCB settings vary.
Check and set water hardness (dH) and electrolysis settings in
service modes (n.181 and n.182) or in the ActiveAqua view.
Reset the counters for prefilter replacement and annual
maintenance.
E83.2
ActiveAqua electrolysis chamber asymmetrical current error.
E83.3
ActiveAqua electrolysis chamber high current error.
E83.4
ActiveAqua electrolysis chamber low current error.
E83.5
ActiveAqua RT clock malfunction.
E83.6
ActiveAqua water flow high.
E83.7
ActiveAqua level sensors missing.
Level sensors are not connected.
E83.8
ActiveAqua electrolysis chamber very high current.
E83.9
ActiveAqua water container fill timeout.
ActiveAqua can not fill the water container. Ensure that
nothing blocks the inlet water (stuck inlet water filter or flow
reducer inside the main water block) and that the water
container does not leak.
E84.1
ActiveAqua water container fill self-test fail.
ActiveAqua can not fill the water container. Ensure that
nothing blocks the inlet water.
E84.2
ActiveAqua electrolysis chamber self-test fail.
Electrolysis chamber current not detected in ActiveAqua self-
test. Check the chamber connection.
31 Help and error messages
User's manual
Planmeca Compact i 275
Содержание Compact i Touch v2
Страница 1: ...PlanmecaCompact iTouchv2 user s manual EN 10033022 ...
Страница 10: ...Table of contents Planmeca Compact i User s manual ...
Страница 27: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 17 ...
Страница 306: ...34 Technical information 296 Planmeca Compact i User s manual ...
Страница 307: ...34 3 5 Side delivery with hanging tube instruments 34 Technical information User s manual Planmeca Compact i 297 ...
Страница 308: ...34 Technical information 298 Planmeca Compact i User s manual ...
Страница 309: ...34 3 6 Cart 34 Technical information User s manual Planmeca Compact i 299 ...
Страница 313: ......