Help messages
HELP
CODE
MEANING
ACTION OR EXPLANATION
H 13
Instrument cannot run because of patient
security violation.
Return the foot control pedal once to the
middle position to start the instrument.
Instruments do not start if the standard foot
control pedal is already pushed to the right
or left when the instrument is picked up from
the instrument console.
H 14
This dental instrument is not recognised
(new type of instrument).
To be able to operate this new instrument
the MCB software must be updated.
Contact your Planmeca dealer.
H 14.1 Instrument presets reset to defaults.
Instrument presets are faulty. Presets are
reset to defaults.
H 15
This button has no function at the moment.
Instruments: Pick up a dental instrument
before attempting to alter its settings; or if
the instrument is selected, this function is
not allowed with this instrument. Other
buttons: the unit does not have this function.
H 16
Operating light is dimmed and
polymerisation light can be used.
This feature enables easier work with
composite materials, minimising the risk of
operating light induced curing. This feature
is triggered by picking up and returning the
polymerisation light to its holder without
starting it. The operating light returns to
normal at the same instance any control is
touched or another instrument selected. The
feature can be disabled.
Contact your Planmeca dealer.
H 17
No syringe connected to instrument
console.
The console syringe must be re-connected,
otherwise there is an air-leak from the
syringe QC whenever an instrument or
device that requires air pressure is used.
H 18
No instrument hose connected to this
instrument place.
The instrument is picked up, but there is no
hose connected at this location. Check that
all hoses go to the right instrument arms.
H 19
No syringe connected to connector located
on cuspidor.
The assistant's syringe must be re-
connected, otherwise there is an air-leak
from the assistant's syringe QC whenever
an instrument or device that requires air
pressure is used.
Contact your Planmeca dealer.
H 20
Wrong instrument is connected to cuspidor
syringe connector or assistant instrument
configuration is wrong.
Contact your Planmeca dealer.
H 21
Flow of both spray air and spray water are
programmed to zero (no flow).
Contact your Planmeca dealer.
H 22
Main air or water valve is set to be always
closed in SM 12 (air) /13 (water).
This is a reminder that either of the main
valves has been set to be always closed.
The setting may be changed in service
modes SM 12 (air) and SM 13 (water).
31 Help and error messages
User's manual
Planmeca Compact i 257
Содержание Compact i Touch v2
Страница 1: ...PlanmecaCompact iTouchv2 user s manual EN 10033022 ...
Страница 10: ...Table of contents Planmeca Compact i User s manual ...
Страница 27: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 17 ...
Страница 306: ...34 Technical information 296 Planmeca Compact i User s manual ...
Страница 307: ...34 3 5 Side delivery with hanging tube instruments 34 Technical information User s manual Planmeca Compact i 297 ...
Страница 308: ...34 Technical information 298 Planmeca Compact i User s manual ...
Страница 309: ...34 3 6 Cart 34 Technical information User s manual Planmeca Compact i 299 ...
Страница 313: ......