After interrupting the cleaning program you can start it again by filling the
dosing cubic and pressing the green start button. The program will start
from the beginning.
1. Dosing cubic
2. Suction tube cleaning holder
3. Start button (green)
4. Cancel button (yellow)
9. When the cleaning program is completed, help message H 42 is
displayed. Return the suction tubes to the suction holder and close the
suction cleaning cover.
10. If suction disinfectant is left in the dosing cubic after the cleaning cycle,
clean the cubic under running water.
Results
The dental unit is now ready for normal operation.
NOTE
Before attaching the handpieces back on the suction tubes, wipe the
identification bushings (1) with Planmeca approved surface disinfectant.
27.6 Waterline cleaning
27.6.1 Introduction
Waterlines are a breeding ground for biofilm. Biofilm may contain bacteria
that are harmful to the patient and to the dental team. This is why the
waterlines of the dental unit should be cleaned once a week with a
disinfectant. The solution is left in the unit overnight and the waterlines of the
27 Flushing and cleaning programs
214 Planmeca Compact i
User's manual
Содержание Compact i Touch v2
Страница 1: ...PlanmecaCompact iTouchv2 user s manual EN 10033022 ...
Страница 10: ...Table of contents Planmeca Compact i User s manual ...
Страница 27: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 17 ...
Страница 306: ...34 Technical information 296 Planmeca Compact i User s manual ...
Страница 307: ...34 3 5 Side delivery with hanging tube instruments 34 Technical information User s manual Planmeca Compact i 297 ...
Страница 308: ...34 Technical information 298 Planmeca Compact i User s manual ...
Страница 309: ...34 3 6 Cart 34 Technical information User s manual Planmeca Compact i 299 ...
Страница 313: ......